Canon Medical recalls Infinix DP-i (CAT-870B) catheterization tables that tilt unexpectedly when liquid spills into the tilt console, potentially causing patient falls. Affected models were sold from 2005-2008.
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.
Canon Medical is recalling certain catheterization tables that tilt unexpectedly when liquid spills into the tilt console. This could cause the table to move suddenly, potentially leading to patient falls during medical procedures.
Liquid spilling into the tilt console creates a short circuit that causes the table to tilt without operator control. This unintended movement could lead to patient falls during medical procedures requiring precise positioning.
Healthcare professionals using the Infinix DP-i (CAT-870B) catheterization tables with the Infinix-8000F X-ray system are most at risk. Patients undergoing angiography procedures may be affected if the table tilts unexpectedly.
Check for model number CAT-870B with serial numbers from the list provided (e.g., 99D08X2101, 99D0582039). These tables were sold between May 2005 and October 2008.
Identify the model number on your catheterization table; affected units are CAT-870B with specific serial numbers listed in the recall.
Reach out to Canon Medical for guidance on whether your table is affected and what steps to take next.
The table tilts unexpectedly when liquid spills into the tilt console, causing a short circuit that could lead to patient falls during medical procedures.
CAT-870B models with serial numbers from the list provided (e.g., 99D08X2101, 99D0582039) manufactured between May 2005 and October 2008.
The recall does not require immediate discontinuation; affected users should check their model and serial number and contact Canon Medical for specific guidance.
We’ve drafted a message you can send Canon Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-0360-2020 — Canon Medical Canon Medical Hello, I own a Canon Medical that is covered by recall Z-0360-2020 from Canon Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.