The FDA announced on Feb. 20, 2019, that Luminex Corporation voluntarily recalled the ARIES HSV 1&2 Assay due to partially sealed side cassette assemblies that could decrease assay sensitivity or stability. The agency issued recall number Z-2283-2019 for the device.
How to tell if you have it
- Brand: Luminex
- Product: ARIES HSV 1&2 Assay
- REF: 50-10017
- UDI: 00840487100295
- Lot AA7308 - Immediately
- Lot AA7485 - 2/23/2019
- Lot AA7545 - 3/13/2019
- Lot AA7625 - 4/10/2019
- Lot AA7630 - 3/16/2019
- Lot AA7636 - 4/3/2019
- Lot AA7945 - 4/25/2019
- Lot AA7985 - 5/8/2019
The hazard
GovInfoNow's records show that a portion of side cassette assemblies had back side seals that did not meet design specifications. This defect could result in a decrease in solution volume and decreased sensitivity or stability of the assay.
Context
Luminex Corporation has 22 other recalls on record since 2018-05-08. Other recent recalls of similar products include Luminex VERIGENE Gram Negative Blood Culture Test Recall (2025-04-16), Luminex Verigene Gram Positive Blood Culture Test Recall (2024-06-20), and Luminex VERIGENE BC-GP Utility Tray Recall (2023-11-07). Other recent recalls of similar products include Siemens Healthcare Syva Emit Disopyramide Assay Recall (2026-06-26), AVID Medical Halyard IV Start Kit Recall (2026-07-22), and Siemens Healthcare Syva EMIT 2000 Quinidine Assay Recall (2026-07-13).