Use caution Medical device recall · FDA · Published April 8, 2019

Integra MoniTorr INS1100 CSF Drainage System Recall: Stop Using Due to Breakage Risk

Integra LifeSciences recalls MoniTorr INS1100 CSF Drainage Systems used with Pole Mount System due to reported breakage of transducer mount stopcocks. Affected patients should stop using the device immediately.

🇺🇸 FDA recall #Z-1311-2019Medical deviceIntegra LifeSciences ›
Take action

Contact the company

Recalling firmIntegra LifeSciences Corp.311 Enterprise Dr · Plainsboro, NJ 08536-3344
✎ GovInfoNow plain-language explainer

Integra LifeSciences is recalling MoniTorr INS1100 CSF Drainage Systems used with Pole Mount System because complaints show these stopcocks can break. This breakage could lead to complications in patients needing cerebrospinal fluid drainage. Read more at govinfonow.com/recall/68418-integra-monitorr-ins1100-csf-drainage-system-recall-stop-using-due.

Source Summary by GovInfoNow · govinfonow.com/recall/68418-integra-monitorr-ins1100-csf-drainage-system-recall-stop-using-due · Updated August 20, 2026 · FDA enforcement record: FDA #Z-1311-2019

Cite this page

“Integra MoniTorr INS1100 CSF Drainage System Recall: Stop Using Due to Breakage Risk.” GovInfoNow, updated August 20, 2026. https://govinfonow.com/recall/68418-integra-monitorr-ins1100-csf-drainage-system-recall-stop-using-due. Based on FDA recall #Z-1311-2019 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1311-2019%22&limit=1).

Integra LifeSciences Corp. recalled the MoniTorr INS1100 INS-1100 CSF Drainage System on April 8, 2019, due to stopcock breakage hazards, according to the FDA. The firm initiated this voluntary Class I recall affecting 42,102 units distributed worldwide.

How to tell if you have it

  • Brand: Integra LifeSciences
  • Product: MoniTorr INS1100 INS-1100 CSF Drainage System used with Pole Mount System
  • Quantity: 42,102 units

The hazard

The FDA said the firm identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock. The recall record does not report specific incidents or injuries.

Context

GovInfoNow's records show Integra LifeSciences Corp. has 35 other recalls on record since Aug. 18, 2017. Recent brand recalls include the AURORA Surgiscope System (Feb. 6, 2025), MediHoney Wound Dressing Gel (Jan. 10, 2025), and CODMAN CERTAS Plus Valve (Dec. 16, 2024). Other recent recalls of similar products include the Codman CerebroFlo EVD Catheter (April 10, 2026), Codman CereLink ICP Sensor Basic Kit (Dec. 12, 2025), and Nipro Medical Hemodialysis System Surdial DX (Oct. 27, 2025).

How to tell if you have this one

Check for the MoniTorr INS1100 CSF Drainage System with Pole Mount System. The product is used for cerebrospinal fluid drainage and has been sold to patients needing brain or spinal fluid monitoring.

  • Brand
    Integra LifeSciences
  • Category
    Medical device
FDA recall event · 3 products

All 3 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class I
  • Status
    Terminated
  • Distribution
    Worldwide Distribution - US Nationwide International distribution to Argentina, American Samoa, Australia, Canada, and Guam.
Recall numberProduct, codes & lots
Z-1310-2019MoniTorr 10100 10-100 CSF Drainage System w/Patient Line One Way Valve MoniTorr 10102 10-102 CSF Drainage System w/Patient Line One Way Valve w MoniTorr 10110 10-110 CSF Drainage System used with Pole Mount System MoniTorr 10140 10-140 CSF Drainage System w/Vinyl Measuring Strip MoniTorr 10150 10-150 CSF Drainage System Simple Bag and Line MoniTorr 10110A 10-110A CSF Drainage System w/o Manifold Stopcock MoniTorr SP0017 SP0017 Special EVD 10-110 w/o Y Site Latex Free Sites MoniTorr SP0042 SP0042 Special EVD 10-100 w/o Y Site MoniTorr SP0090 SP0090 Special EVD 10-110 w/o Y Site MoniTorr SP0164 SP0164 Special EVD 10-140 w/Y Site & Stopcock Reverse MoniTorr SP0236 SP0236 INS 1100 WITH NEEDLE AT Y-ACCESS AND NEEDLELESS SI Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use, closed system for the drainage of cerebrospinal fluid (CSF) from ventricles of the brain or the lumbar subarachnoid space to a drainage bag. The system may be used with a pole mounted assembly that allows for simple, quick and accurate alignment with the patient and secure positive or negative pressure level setting. The system also has been designed to provide for ease of patient transport through a compact design and antimicrobial hydrophobic vent feature that resists occlusion. Indications: The MoniTorr ICP system allows for drainage and monitoring of CSF from the lateral ventricles of the brain and the lumbar subarachnoid space in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF inpatients with infected CSF shunts, and to monitor ICP.; All lots
Z-1311-2019MoniTorr INS1100 INS-1100 CSF Drainage System used with Pole Mount System Product Usage: The MoniTorr ICP External CSF Drainage and Monitoring Systems provide a simple to use, closed system for the drainage of cerebrospinal fluid (CSF) from ventricles of the brain or the lumbar subarachnoid space to a drainage bag. The system may be used with a pole mounted assembly that allows for simple, quick and accurate alignment with the patient and secure positive or negative pressure level setting. The system also has been designed to provide for ease of patient transport through a compact design and antimicrobial hydrophobic vent feature that resists occlusion. Indications: The MoniTorr ICP system allows for drainage and monitoring of CSF from the lateral ventricles of the brain and the lumbar subarachnoid space in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF inpatients with infected CSF shunts, and to monitor ICP.; All lots
Z-1312-2019LimiTorr INS9020 INS-9020 LIMITORR VOLUME LIMITING EVD 20 ML LimiTorr INS9020SP1 INS-9020SP1 LIMITORR VOLUME EVD 20ML LimiTorr INS9030 INS-9030 LIMITORR VOLUME LIMITING EVD 30 ML LimiTorr INS9030SP1 INS-9030SP1 LIMITORR VOLUME LIMITING EVD 30 ML Product Usage: The LimiTorr External CSF Drainage and Monitoring Systems provides a closed system for the CSF from the ventricles of the brain or the lumbar subarachnoid space to an external drainage bag. The LimiTorr volume limiting external CSF drainage system was designed to include a volume limiting valve mechanism which reduces the chances for excessive CSF drainage. The burette in the volume limiting external CSF drainage system contains a volume limiting valve which stops drainage when the pre-determined volume (20ml-30ml) is reached. Indications: The LimiTorr system allows for drainage and monitoring of CSF from the lateral ventricles of the brain and the lumbar subarachnoid space in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF in patients with infected CSF shunts, and to monitor ICP.; All lots
Based on the official FDA remedy

What to do — step by step

  1. Stop Using Immediately

    The MoniTorr INS1100 CSF Drainage System with Pole Mount System should be stopped using immediately due to reported breakage risks.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide Distribution - US Nationwide International distribution to Argentina, American Samoa, Australia, Canada, and Guam.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 22 months after the firm started it.

1. Firm started the recall: Apr 8, 201912. FDA classified it Class I: May 20, 201923. Published in FDA’s Enforcement Report: May 29, 201934. FDA terminated the recall: Jan 27, 2021420192021
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1311-2019 (April 8, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1311-2019 to see it in context.
  • Plain-language summary written by qwen3:4b on August 20, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA