The FDA announced on May 22, 2019, that Abbott Laboratories is recalling Hemoglobin A1c Calibrators because a manufacturing defect in some calibrator level 2 bottles may cause biased patient test results. The firm is investigating the root cause of this Class II recall.
How to tell if you have it
- Brand: Abbott
- Product: HEMOGLOBIN A1c CALIBRATORS * REF 4P52-01 * 307261/R03 * S4P5X0 * FOR USE WITH ARCHITECT
- List Number: 4P52-01
- Lot Number: 54582UQ02
- Quantity: 198 Kits
The hazard
The calibrator level 2 bottle may have been manufactured incorrectly for a portion of the lot. While no impact to assay performance was observed between the HbA1c concentration range of 5.33% to 6.87%, a bias of greater than 3% may be seen with patient results outside of this range. This device failure can result in potentially negatively and positively biased patient test results for HbA1c.
Context
GovInfoNow's records show Abbott Laboratories has 33 other recalls on record since May 14, 2012. Recent brand recalls include the November 13, 2025 Alinity ci-series System Control Module correction and the September 18, 2025 Alinity hq Analyzer Software Issue regarding incorrect results when saturation is present. Other recent recalls of similar products include the November 24, 2025 Abbott Diabetes Care FreeStyle Libre 3 Plus Sensor Recall for incorrect low glucose results and the June 8, 2026 Siemens Atellica CH A1c_E Test Recall for false HbA1c results risk.