The FDA announced on May 24, 2019, that Cook Inc. is recalling the Advance Enforcer 35 Focal Force PTA Balloon Catheter due to balloons bursting below rated pressure. This Class I recall affects 73 units distributed worldwide.
How to tell if you have it
- Brand: Cook Inc.
- Product: Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/135cm
- Catalog Number: ASB5-35-135-6-4
- REF Number: G35257
- Lot Numbers: 9338194, 9234423, 9278982, 9209468, 9248603, 9320429
The hazard
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.
Context
GovInfoNow's records show Cook Inc. has 38 other recalls on record since July 31, 2017. Other recent recalls of similar products include Cook Flexor Check-Flo Introducer Ansel Modification (Nov. 12, 2021), Cook Skinny Needle with Chiba Tip Recall (Oct. 13, 2021), and Cook Transseptal Needle with Catheter Recall (Oct. 8, 2021). Other recent recalls of similar products include LeMaitre TufTex Catheter Recall (Aug. 28, 2026), Cook Aurous Sizing Catheters Recall (April 2, 2026), and Microvention Bobby Balloon Guide Catheter Recall (April 30, 2026).