Becton Dickinson recalls BD Vacutainer Fluoride Tubes (Part 367001) for blood alcohol tests due to potential inaccurate results if stored over 2 days. Affected lots may need replacement to ensure accurate blood alcohol measurements.
Lot has been confirmed to have no additive within the tube. Sample processed without the preservative (additive) in the tube, testing has yielded reliable results if the samples were stored at room temperature for no longer than two days. If the sample was stored for more than 2 days, the result for blood alcohol determination might not be accurate (either falsely low or falsely high)
Becton Dickinson is recalling Fluoride Tubes used for blood alcohol tests because they might give inaccurate results if stored longer than two days at room temperature. This could lead to incorrect blood alcohol readings, which is important for legal testing.
The tubes lack a preservative additive that normally keeps samples stable. If stored at room temperature for more than two days, the blood alcohol test results may be falsely low or falsely high, leading to incorrect legal conclusions.
People who use BD Vacutainer Fluoride Tubes (Part Number 367001) for blood alcohol testing are affected. Those who store samples longer than two days at room temperature are at higher risk of inaccurate results.
Check for the Part Number 367001 on the tube. The lot number 8187663 and expiration date 2020/7/31 are also listed on the tube's unique identifier (UDI).
If the sample has been stored at room temperature for more than two days, the test result may be inaccurate.
The tubes lack a preservative additive that keeps samples stable. If stored at room temperature for more than two days, blood alcohol test results may be inaccurate.
Look for Part Number 367001 on the tube. The lot number 8187663 and expiration date 2020/7/31 are also listed on the tube's unique identifier (UDI).
No, the remedy is not stated in the recall notice. You can continue using the tubes as long as samples are stored at room temperature for no longer than two days.
We’ve drafted a message you can send Becton Dickinson to request your refund or repair — edit it as you like.
Subject: Recall Z-2279-2019 — Becton Dickinson Becton Dickinson Hello, I own a Becton Dickinson that is covered by recall Z-2279-2019 from Becton Dickinson. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.