Skip to content
Use caution Medical device recall · FDA · Published June 26, 2019

Deerfield Imaging HFD200 Head Fixation Device Recall: Risk of Improper Skull Pinning

Deerfield Imaging recalls HFD200 MR/X-ray head fixation devices due to potential improper skull pinning from drive knob force indicator failure. Affected units have specific serial numbers.

🇺🇸 FDA recall #Z-0163-2020Medical deviceDeerfield Imaging ›
Take action

Contact the company

Recalling firmDeerfield Imaging, Inc.5101 Shady Oak Rd S · Minnetonka, MN 55343-4100
✎ GovInfoNow plain-language explainer

Deerfield Imaging is recalling its HFD200 head fixation device because the drive knob force indicator can break under over-torqueing, leading to either insufficient or excessive force on the patient's skull. This could cause injury during medical procedures if not properly monitored. Read more at govinfonow.com/recall/67581-deerfield-imaging-hfd200-head-fixation-device-recall-risk-of-improper.

Source Summary by GovInfoNow · govinfonow.com/recall/67581-deerfield-imaging-hfd200-head-fixation-device-recall-risk-of-improper · Updated August 20, 2026 · FDA enforcement record: FDA #Z-0163-2020

Cite this page

“Deerfield Imaging HFD200 Head Fixation Device Recall: Risk of Improper Skull Pinning.” GovInfoNow, updated August 20, 2026. https://govinfonow.com/recall/67581-deerfield-imaging-hfd200-head-fixation-device-recall-risk-of-improper. Based on FDA recall #Z-0163-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0163-2020%22&limit=1).

Deerfield Imaging, Inc. voluntarily recalled its IMRIS MR/X-ray head fixation device, Model HFD200, on June 26, 2019, according to the Food and Drug Administration (FDA). The recall addresses a hazard where breaking of the drive knob force indicator may result in insufficient or excessive skull pinning force. Three units were distributed in Tennessee, New Hampshire, and Florida.

How to tell if you have it

  • Brand: Deerfield Imaging
  • Product: IMRIS MR/X-ray head fixation device
  • Model: HFD200
  • Units affected: 3
  • Distribution locations: TN, NH, FL

The hazard

The FDA stated that breaking of the HFD200 Drive Knob force indicator due to over-torqueing may result in clinical staff unknowingly pinning the patient's skull with insufficient or excessive force. The hazard also includes the possibility that staff identify the issue but cannot continue with the procedure.

Context

GovInfoNow's records show Deerfield Imaging, Inc. has had eight other recalls on record since Feb. 20, 2018. Other recent recalls of similar products include HFD100 Head Fixation Device Recall (Dec. 8, 2025), Olympus HX-400U-30 Ligating Device Recall (Oct. 30, 2025), and Hologic 3D Mammography Systems Recall (April 8, 2026).

How to tell if you have this one

Check for the HFD200 model with specific serial numbers: 10002481, 10003305, 10002351, 10005723, 10004247, 10002448. These units were sold with the drive knob force indicator that can fail.

  • Brand
    Deerfield Imaging
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Serial Numbers

    Verify if your HFD200 device has one of the listed serial numbers: 10002481, 10003305, 10002351, 10005723, 10004247, 10002448.

Where it went and when

Where it was distributed

Distributed in 3 states.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: yesNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: yesTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: yesFL

FDA distribution: “TN, NH, FL” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Florida, New Hampshire, Tennessee.

Recall timeline

FDA terminated this recall 16 months after the firm started it.

1. Firm started the recall: Jun 26, 201912. FDA classified it Class II: Oct 18, 201923. Published in FDA’s Enforcement Report: Oct 30, 201934. FDA terminated the recall: Oct 29, 2020420192020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0163-2020 (June 26, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0163-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 20, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA