Degania Silicone recalls 15FR channel drains with risk of breaking during removal, requiring additional surgery. Product number JP-218: P1841752, P1852268, P1845823.
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
Degania Silicone recalls Cardinal Health 15FR channel drains that may break during removal, causing extra surgery. This affects specific product numbers sold in certain lots.
The drain channel may break at the connection point between the clear tube and white profile part during removal. This requires additional surgery to remove the broken piece from the patient's body.
Patients using the recalled channel drains for drainage procedures. Those with the specific product numbers and lots are most at risk.
Check for product number JP-2189 on the drain. Look for lot codes P1841752, P1852268, or P1845823. The drain has a trocar and full flutes.
Look for product number JP-2189 on the drain.
Check if your drain has lot codes P1841752, P1852268, or P1845823.
The drain may break during removal at the connection point between the clear tube and white profile part, requiring additional surgery to remove the broken piece.
Check the product number and lot codes to see if it's recalled. If it is, contact Cardinal Health for instructions.
Yes, breaking during removal can require additional surgery, but the remedy is not stated in the recall record.
We’ve drafted a message you can send Degania Silicone to request your refund or repair — edit it as you like.
Subject: Recall Z-2262-2019 — Degania Silicone Degania Silicone Hello, I own a Degania Silicone that is covered by recall Z-2262-2019 from Degania Silicone. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.