Degania Silicone recalls Cardinal Health Jackson-Pratt Channel Drains 19FR model with risk of breaking during removal, requiring additional surgery. Affected units have specific lot codes.
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
Degania Silicone recalls Cardinal Health Jackson-Pratt Channel Drains 1:19FR model that may break during removal, causing the need for additional surgery. This is a serious risk for patients using the drain.
The drain's round fluted channel may break at the connection point between the clear tube and white profile part during removal. This breakage can require additional surgery to remove the broken piece from the patient's body.
Patients using the Cardinal Health Jackson-Pratt Channel Drain 19FR model with trocar are affected. Those who have the drain in place and are undergoing surgery are at highest risk.
Check for the model number '19FR' on the drain. The product was sold with lot codes P1861538, P1865416, and P1903570.
Immediately stop using the drain if it is broken or causing pain.
The drain may break during removal, requiring additional surgery to remove the broken piece from the patient's body.
The affected model is Cardinal Health Jackson-Pratt Channel Drain 19FR with trocar.
Lot codes P1861538, P1865416, and P1903570 are affected.
We’ve drafted a message you can send Degania Silicone to request your refund or repair — edit it as you like.
Subject: Recall Z-2264-2019 — Degania Silicone Degania Silicone Hello, I own a Degania Silicone that is covered by recall Z-2264-2019 from Degania Silicone. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.