Degania Silicone recalls 15FR Cardinal Health channel drains that may break during removal, requiring additional surgery. Check lot codes 1900956 or P190:00955 for affected units.
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
Degania Silicone recalls Cardinal Health Jackson-Pratt Channel Drains 15FR (model JP-2223) that can break during removal, potentially needing extra surgery. This risk affects patients using these drains post-surgery.
The drain's round fluted channel may break when being removed from the body, especially at the connection point between the clear tube and white profile. This breakage can require additional surgery to remove the broken piece.
Patients who recently had surgery requiring a channel drain are most at risk. The product is used for draining body cavities after medical procedures.
Check the product label for lot codes 1900956 or P1900955. The drain is a 15FR Cardinal Health Jackson-Pratt Channel Drain with model number JP-2223.
Look for lot codes 1900956 or P1900955 on the drain's packaging or label.
The drain may break during removal, requiring additional surgery to remove the broken piece.
Check the product label for lot codes 1900956 or P1900955.
The recall does not require immediate stop use; check the lot code to see if your drain is affected.
We’ve drafted a message you can send Degania Silicone to request your refund or repair — edit it as you like.
Subject: Recall Z-2260-2019 — Degania Silicone Degania Silicone Hello, I own a Degania Silicone that is covered by recall Z-2260-2019 from Degania Silicone. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.