Skip to content
Use caution Medical device recall · FDA · Published July 3, 2019

Elekta Monaco RTP Software 5.10/5.11: Incorrect Monitor Units Scaling Risk

Elekta recalls Monaco RTP software versions 5.10 and 5.11 due to potential incorrect monitor unit scaling during radiation treatment planning. Affected units may require software update to prevent treatment errors.

🇺🇸 FDA recall #Z-2091-2019Medical deviceElekta ›
Take action

Contact the company

Recalling firmElekta Inc13723 Riverport Dr Ste 100 · Maryland Heights, MO 63043-4819
✎ GovInfoNow plain-language explainer

What happened, in plain English

Elekta is recalling its Monaco RTP software versions 5.10 and 5.11 for radiation treatment planning systems. This recall occurs because changing treatment parameters can cause incorrect monitor unit calculations, potentially leading to unsafe radiation doses. Read more at govinfonow.com/recall/67503-elekta-monaco-rtp-software-5-10-5-11-incorrect-monitor.

Source Summary by GovInfoNow · govinfonow.com/recall/67503-elekta-monaco-rtp-software-5-10-5-11-incorrect-monitor · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2091-2019

Cite this page

“Elekta Monaco RTP Software 5.10/5.11: Incorrect Monitor Units Scaling Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/67503-elekta-monaco-rtp-software-5-10-5-11-incorrect-monitor. Based on FDA recall #Z-2091-2019 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2091-2019%22&limit=1).

Why it’s dangerous

When creating 3D treatment plans using specific modes, changing the number of radiation fractions, rescaling the plan, and then adjusting the wedge angle can cause the software to calculate incorrect monitor units. This error may lead to improper radiation doses being delivered to patients.

Who’s affected

Radiation therapy professionals using Elekta Monaco RTP systems with software versions 5.10 or 5.11 are most at risk. Patients receiving treatments with these systems could be affected if the software error occurs.

How to tell if you have this one

Check if your Monaco RTP system has software version 5.10 or 5.11. The recall applies to all units installed with these software versions.

  • Brand
    Elekta
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Update Software

    Install the corrected software version from Elekta to fix the scaling issue.

How this recall unfolded

Recall timeline

FDA published this recall 35 days after the firm started it.

1. Firm started the recall: Jul 3, 201912. FDA classified it Class II: Jul 27, 201923. Published in FDA’s Enforcement Report: Aug 7, 201932019
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2091-2019 (July 3, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2091-2019 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA