Skip to content
Use caution Medical device recall · FDA · Published July 8, 2019

CooperSurgical TransWarmer Infant Mattress Recall: Risk of Skin Burns from Incubator Use

CooperSurgical recalls TransWarmer Warming Infant Transport Mattress Model 20421 due to risk of skin burns if used with incubators. Some units lack updated safety instructions.

🇺🇸 FDA recall #Z-2364-2019Medical deviceCooperSurgical ›
Take action

Contact the company

Recalling firmCooperSurgical, Inc.95 Corporate Dr · Trumbull, CT 06611-1350
✎ GovInfoNow plain-language explainer

CooperSurgical is recalling their TransWarmer Warming Infant Transport Mattress (Model 20421) because using it with other heat sources like incubators can cause serious skin burns. This recall affects hospitals and caregivers who use the product for infant transport. Read more at govinfonow.com/recall/67478-coopersurgical-transwarmer-infant-mattress-recall-risk-of-skin-burns-from.

Source Summary by GovInfoNow · govinfonow.com/recall/67478-coopersurgical-transwarmer-infant-mattress-recall-risk-of-skin-burns-from · Updated August 20, 2026 · FDA enforcement record: FDA #Z-2364-2019

Cite this page

“CooperSurgical TransWarmer Infant Mattress Recall: Risk of Skin Burns from Incubator Use.” GovInfoNow, updated August 20, 2026. https://govinfonow.com/recall/67478-coopersurgical-transwarmer-infant-mattress-recall-risk-of-skin-burns-from. Based on FDA recall #Z-2364-2019 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2364-2019%22&limit=1).

CooperSurgical, Inc. recalled the TransWarmer Warming Infant Transport Mattress, Model no. 20421, because some units lack an updated instruction manual that prohibits use with other heat-producing devices, which could cause skin burns, according to the FDA. The agency announced the recall on July 8, 2019.

How to tell if you have it

  • Brand: CooperSurgical
  • Product: TransWarmer Warming Infant Transport Mattress
  • Model no.: 20421
  • Lot codes: IJ780, IJ782, IJ785, IJ789, IJ794, IJ796, IJ804, IJ807, IJ817, IJ824, IJ832, IJ834, IJ841, IJ853, IJ865, IJ874, IJ878, IJ884, IJ886, IJ892, IJ895, IJ897, IJ900, IJ902, IJ909, IJ917, IJ920, IJ924, IJ933, IJ946, IJ948, IJ957, IJ962, IJ973, IJ976, IJ982, IJ985, IJ999, IK007, IK009, IK015, IK028, IK038, IK041, IK053, IK060, IK066, IK095, IK102, IK107, IK127, IK135, IK147, IK162, IK164, IK171, IK178, IK184, IK201, IK214, IK238, IK252, IK258, IK265, IK270, IK276, IK298, IK318, IK322, IK330, IK332, IK339, IK353, IK361, IK378, IK384, & IK390

The hazard

The firm updated the instruction manual in April 2019 to clarify that using the TransWarmer Mattress with other heat-producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. Some products do not contain the updated instruction manual.

Context

GovInfoNow's records show CooperSurgical, Inc. has 37 other recalls on record since October 6, 2017. Other recent recalls of similar products include INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP (August 11, 2026), Endosee System Convenience Kit (June 11, 2025), and K-Systems T47 Warming Plate (November 15, 2024).

How to tell if you have this one

Check for the model number 20421 on the product. The recall applies to units sold with older safety instructions that don't include the April 2019 update.

  • Brand
    CooperSurgical
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Safety Instructions

    Look for the updated safety instructions from April 2019 that prohibit using the mattress with incubators or other heat sources.

  2. Stop Using with Incubators

    If you have the mattress with older instructions, do not use it with incubators or other heat-producing devices.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 11 months after the firm started it.

1. Firm started the recall: Jul 8, 201912. FDA classified it Class II: Aug 23, 201923. Published in FDA’s Enforcement Report: Sep 4, 201934. FDA terminated the recall: May 28, 2020420192020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2364-2019 (July 8, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2364-2019 to see it in context.
  • Plain-language summary written by qwen3:4b on August 20, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA