Use caution Medical device recall · FDA · Published July 31, 2019

Haag-Streit HS Hi-R NEO 900A NIR Software Error Recall: Risk of Incorrect Surgical

Haag-Streit recalls specific models of surgical microscopes with software errors that may cause incorrect surgical measurements during procedures. Affected units include certain FS 2-21, FS 2-25, and FS 3-45 floor stands with specific serial numbers and software versions.

🇺🇸 FDA recall #Z-2484-2019Medical deviceHaag-Streit ›
Take action

Contact the company

Recalling firmHaag-Streit USA Inc3535 Kings Mills Rd · Mason, OH 45040-2303
✎ GovInfoNow plain-language explainer

What happened, in plain English

Haag-Streit is recalling specific models of surgical microscopes that have a software error in the central control unit. This error could lead to incorrect surgical measurements during procedures, which might affect patient safety. Read more at govinfonow.com/recall/67270-haag-streit-hs-hi-r-neo-900a-nir-software-error.

Source Summary by GovInfoNow · govinfonow.com/recall/67270-haag-streit-hs-hi-r-neo-900a-nir-software-error · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2484-2019

Cite this page

“Haag-Streit HS Hi-R NEO 900A NIR Software Error Recall: Risk of Incorrect Surgical.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/67270-haag-streit-hs-hi-r-neo-900a-nir-software-error. Based on FDA recall #Z-2484-2019 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2484-2019%22&limit=1).

Why it’s dangerous

The software error in the central control unit can cause incorrect surgical measurements when used with specific floor stands (FS 2-21, FS 2-25, FS 3-45) and software versions 10.1.xx to 11.3.xx. This may lead to unintended surgical outcomes if the error affects critical measurements during procedures.

Who’s affected

Surgical professionals using Haag-Streit HS Hi-R NEO 900A NIR microscopes with specific floor stands and software versions are most at risk. Patients undergoing procedures with these affected devices may experience inaccurate measurements.

How to tell if you have this one

Check if your microscope has the model HS Hi-R NEO 900A NIR with the floor stand model FS 2-21, FS 2-25, or FS 3-45. Look for serial numbers and software versions listed in the recall notice: serial numbers from 101 to 360 or 361, and software versions 10.1.xx to 11.3.xx.

  • Brand
    Haag-Streit
  • Category
    Medical device
FDA recall event · 3 products

All 3 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    US in the state of OH
Recall numberProduct, codes & lotsQuantity
Z-2482-2019HS Hi-R NEO 900 Reference # 657 820 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating microscopes are qualified for further applications, in particular ENT- and reconstructive microsurgery are possible, because their line of sight can be tilted or adjusted horizontally. The HS ALLEGRA 590 is particular because of its multidisciplinary application possibilities.; The Error only occurs in combination with the following floor stands FS 2-21, FS 2-25, FS 3-45 with the software REF 615 586 in the versions 10.1.xx to 11.3.xx 1. Product: FS 2-21, Reference # 615 510 Serial # 101 to 360, SW-Version: 10.1.xx to 11.3.xx 2. Product: FS 2-21, Reference # 615 510, serial # from 361 3. Product: FS 2-25, Reference # 615 550, Serial # 101 to 376, SW Version: 10.1.xx to 11.3.xx 4. Product: FS 2-25, Reference # 615 550, Serial # from 377 4. Product: FS 3-45, Reference # 615 705, Serial #from 1010
Z-2483-2019HS Hi-R NEO 900A Reference # 657 821 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating microscopes are qualified for further applications, in particular ENT- and reconstructive microsurgery are possible, because their line of sight can be tilted or adjusted horizontally. The HS ALLEGRA 590 is particular because of its multidisciplinary application possibilities.; The Error only occurs in combination with the following floor stands FS 2-21, FS 2-25, FS 3-45 with the software REF 615 586 in the versions 10.1.xx to 11.3.xx 1. Product: FS 2-21, Reference # 615 510 Serial # 101 to 360, SW-Version: 10.1.xx to 11.3.xx 2. Product: FS 2-21, Reference # 615 510, serial # from 361 3. Product: FS 2-25, Reference # 615 550, Serial # 101 to 376, SW Version: 10.1.xx to 11.3.xx 4. Product: FS 2-25, Reference # 615 550, Serial # from 377 4. Product: FS 3-45, Reference # 615 705, Serial #from 1010
Z-2484-2019HS Hi-R NEO 900A NIR Reference # 657 822 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating microscopes are qualified for further applications, in particular ENT- and reconstructive microsurgery are possible, because their line of sight can be tilted or adjusted horizontally. The HS ALLEGRA 590 is particular because of its multidisciplinary application possibilities.; The Error only occurs in combination with the following floor stands FS 2-21, FS 2-25, FS 3-45 with the software REF 615 586 in the versions 10.1.xx to 11.3.xx 1. Product: FS 2-21, Reference # 615 510 Serial # 101 to 360, SW-Version: 10.1.xx to 11.3.xx 2. Product: FS 2-21, Reference # 615 510, serial # from 361 3. Product: FS 2-25, Reference # 615 550, Serial # 101 to 376, SW Version: 10.1.xx to 11.3.xx 4. Product: FS 2-25, Reference # 615 550, Serial # from 377 4. Product: FS 3-45, Reference # 615 705, Serial #from 1011
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your microscope's software version by checking the device settings or documentation. The affected versions are 10.1.xx to 11.3.xx.

  2. Contact Manufacturer

    Reach out to Haag-Streit USA Inc for guidance on whether your specific unit is affected and what steps to take next.

Where it went and when

Where it was distributed

Distributed in Ohio.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: yesOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: not listedFL

FDA distribution: “US in the state of OH” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Ohio.

Recall timeline

FDA terminated this recall 9 months after the firm started it.

1. Firm started the recall: Jul 31, 201912. FDA classified it Class II: Sep 5, 201923. Published in FDA’s Enforcement Report: Sep 11, 201934. FDA terminated the recall: Apr 16, 2020420192020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2484-2019 (July 31, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2484-2019 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA