The FDA announced on 2019-08-22 that ARJOHUNTLEIGH POLSKA Sp. z.o.o. is recalling Concerto & Basic Shower Trollys due to a potential patient fall hazard. The firm initiated this voluntary Class II recall after discovering mechanical failures in the safety catches.
How to tell if you have it
- Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102
- Concerto & Basic Shower Trolly, Model Number BAB1007-01 (UDI: (01)05055982796255)
- Concerto & Basic Shower Trolly, Model Number BAB1101-01 (UDI: (01)05055982787949)
- Concerto & Basic Shower Trolly, Model Number BAB1102-01 (UDI: (01)05055982796156)
- Concerto & Basic Shower Trolly, Model Number BAB1103-01 (UDI: (01)05055982796163)
- Concerto & Basic Shower Trolly, Model Number BAB1107-01 (UDI: (01)05055982796187)
- Concerto & Basic Shower Trolly, Model Number BAB2000-01 (UDI: (01)05055982788106)
- Concerto & Basic Shower Trolly, Model Number BAB2101-01 (UDI: (01)05055982788007)
- Concerto & Basic Shower Trolly, Model Number BAB2102-01 (UDI: (01)05055982796194)
- Concerto & Basic Shower Trolly, Model Number BAB3000-01 (UDI: (01)05055982788120)
- Concerto & Basic Shower Trolly, Model Number BAB3101-01 (UDI: (01)05055982788021)
- Concerto & Basic Shower Trolly, Model Number BAB3107-01 (UDI: (01)05055982796248)
- Concerto & Basic Shower Trolly, Model Number BAB5000-01 (UDI: (01)05055982788144)
- Concert & Basic Shower trolleys Serial Numbers: P0512263 to P0551621, P0279164, SEE0643339, SEE0605124,P0249502, SEE045073, SEE0520032, 1106000338, SEE0348065, P0422951, P0232526
- Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031
The hazard
The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure, specifically a break in one of the arms of the safety catch. This failure may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall. The patient falling could lead to various types of injuries.
Context
GovInfoNow's records show ARJOHUNTLEIGH POLSKA Sp. z.o.o. has 6 other recalls on record since 2018-10-22. Other recent recalls of similar products include Arjo Tenor Patient Lift Recall (2026-03-10), Arjo medical beds with IndiGo Drive wheels recall (2023-08-29), and Sara Plus Active Floor Lift Recall (2022-04-05). Other recent recalls of similar products include Olymp ShockPulse-SE Lithotripsy System Generator Recall (2025-11-14), Kiwi Omni Vacuum Delivery System Recall (2026-03-09), and Accuray Cyberknife System Recall (2026-06-16).