Philips recalls IntelliVue 1.4 GHz Remote Antennas used with MX40 monitors that have inadequate cable ratings for certain installations. Affected units shipped since May 2019 with 'CM' marked coax data cable. Serial numbers listed in the recall.
Remote Antenna Cable Used with Philips MX4O Monitors Does Not Have the Required Rating for Some Installations
Philips is recalling a 74-foot IntelliVue 1.4 GHz Remote Antenna used with MX40 monitors because the cable lacks the required rating for some installations. This could pose a risk if the cable fails under electrical load. Affected units were shipped since May 2019 with 'CM' marked coax data cable.
The remote antenna cable used with Philips MX40 monitors does not have the required electrical rating for certain installations. This could lead to potential failure under electrical load, but the hazard is not described as causing immediate injury or death in normal use conditions.
Users of Philips IntelliVue MX40 Patient Worn Monitors in the USA who have the affected Remote Antenna (PN 867151) shipped since May 2019 with 'CM' marked coax data cable. Most at risk are healthcare professionals using these monitors in clinical settings.
Check if your IntelliVue Remote Antenna (PN 867151) has a 'CM' marked coax data cable and was shipped since May 2019. Serial numbers listed in the recall include many combinations starting with DK8200 or DK7490.
Look for the serial number on the Remote Antenna (PN 867151) to see if it matches the list of affected serial numbers provided in the recall.
If your serial number is on the affected list, contact Philips customer service for instructions on how to return or replace the antenna.
The affected antenna is the IntelliVue 1.4 GHz Remote Antenna (PN 867151) used with Philips MX40 Patient Worn Monitors.
These units were shipped since May 2019.
Check if your antenna has a 'CM' marked coax data cable and a serial number listed in the provided list of affected serial numbers.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-2547-2019 — Philips Philips Hello, I own a Philips that is covered by recall Z-2547-2019 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.