Spacelabs Xhibit Central Model 96102: Audio alarm loss after power failure

Spacelabs Healthcare recalls Xhibit Central Model 96102 software versions 1.3.1-1.3.4 due to potential audio alarm loss after power failure or cable disconnection, which could delay recognizing patient emergencies.

🇺🇸 FDA recall #Z-2261-2020Medical deviceSpacelabs Healthcare ›
Take action

Contact the company

Recalling firmSpacelabs Healthcare, Inc.35301 Se Center St · Snoqualmie, WA 98065-9216
✎ GovInfoNow plain-language explainer

Spacelabs Healthcare is recalling Xhibit Central Model 96102 software versions 1.3.1 through 1.3.4. This device may lose audio alarms after power failures or cable disconnections, potentially delaying recognition of critical patient conditions. Read more at govinfonow.com/recall/67084-spacelabs-xhibit-central-model-96102-audio-alarm-loss-after-power.

Source Summary by GovInfoNow · govinfonow.com/recall/67084-spacelabs-xhibit-central-model-96102-audio-alarm-loss-after-power · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2261-2020

Cite this page

“Spacelabs Xhibit Central Model 96102: Audio alarm loss after power failure.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/67084-spacelabs-xhibit-central-model-96102-audio-alarm-loss-after-power. Based on FDA recall #Z-2261-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2261-2020%22&limit=1).

The FDA announced on Aug. 30, 2019, that Spacelabs Healthcare, Inc. is recalling Xhibit Central, Model No. 96102, due to a hazard where loss of audio alarms after power failure or cable disconnection could delay recognition of alarm conditions for telemetry patients.

How to tell if you have it

  • Brand: Spacelabs Healthcare
  • Product: Xhibit Central
  • Model No.: 96102
  • Software Versions: 1.3.1, 1.3.2, 1.3.3 and 1.3.4
  • UDI: 10841522100345
  • Quantity: 2197 units
  • Distribution: US Nationwide (including 40 states and Puerto Rico), Canada, and multiple other countries

The hazard

The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.

Context

GovInfoNow's records show Spacelabs Healthcare, Inc. has 10 other recalls on record since Jan. 17, 2018. Other recent recalls of similar products include Philips IntelliVue MX70: Monitor May Not Alarm - Recall (Oct. 31, 2025), Philips IntelliVue MX50: Monitor May Not Alarm - Recall (Oct. 31, 2025), and Philips IntelliVue MX550 Patient Monitor Recall: No Alarms May Cause Missed Alerts (Oct. 31, 2025).

How to tell if you have this one

Check for Xhibit Central Model 96102 with software versions 1.3.1, 1.3.2, 1.3.3, or 1.3.4. The device is used for central monitoring of adult, pediatric, and neonatal patient data in healthcare settings.

  • Brand
    Spacelabs Healthcare
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your Xhibit Central software version by checking the device's display or documentation.

  2. Update Software

    Contact Spacelabs Healthcare to update to a newer software version that resolves this issue.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Worldwide distribution - US Nationwide distribution including in the 40 states and Puerto Rico, and multiple countries including Canada.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 4.6 years after the firm started it.

1. Firm started the recall: Aug 30, 201912. FDA classified it Class II: Jun 4, 202023. Published in FDA’s Enforcement Report: Jun 10, 202034. FDA terminated the recall: Apr 18, 2024420192024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2261-2020 (August 30, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2261-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA