Skip to content
Use caution Medical device recall · FDA · Published August 30, 2019

Philips Transducer S8-3t: Temperature Measurement Risk in TEE Ultrasound

Philips recalls S8-3t transducers for potential patient injury due to incorrect temperature measurement during transesophageal echocardiography (TEE) scans.

🇺🇸 FDA recall #Z-1843-2020Medical devicePhilips ›
Take action

Contact the company

Recalling firmPhilips Ultrasound Inc22100 Bothell Everett Hwy · Bothell, WA 98021-8431
✎ GovInfoNow plain-language explainer

Philips is recalling some S8-3t transducers used with their ultrasound systems because they may not properly measure temperature during use. This could cause the transducer to overheat and injure patients during TEE scans. Read more at govinfonow.com/recall/67082-philips-transducer-s8-3t-temperature-measurement-risk-in-tee-ultrasound.

Source Summary by GovInfoNow · govinfonow.com/recall/67082-philips-transducer-s8-3t-temperature-measurement-risk-in-tee-ultrasound · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1843-2020

Cite this page

“Philips Transducer S8-3t: Temperature Measurement Risk in TEE Ultrasound.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/67082-philips-transducer-s8-3t-temperature-measurement-risk-in-tee-ultrasound. Based on FDA recall #Z-1843-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1843-2020%22&limit=1).

The FDA announced on 2019-08-30 that Philips Ultrasound Inc is recalling Transducer S8-3t devices because they may fail to measure temperature correctly, posing a risk of patient injury. The agency stated the recall addresses transducers not properly programmed during manufacturing.

How to tell if you have it

  • Brand: Philips
  • Product: Transducer S8-3t
  • UDI: (01)00884838073524
  • REF: 989605431171
  • UPC: 00884838073524
  • Compatible systems: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi

The hazard

According to the FDA, some Transesophageal (TEE) transducers were not properly programmed during manufacturing, causing the system to be unable to correctly measure the temperature on the distal tip. This issue may result in affected transducers not initiating auto-cool during use, which could result in patient injury.

Context

GovInfoNow's records show Philips Ultrasound, Inc has had 14 other recalls since 2019-04-29. Recent similar recalls include the Philips L12-5 Ultrasound Transducer Recall on 2025-09-05 and the Philips SCNHD LA L7-... Transducer Probe Recall on 2025-07-03, both addressing useful life labeling. Other recent recalls of similar products include the Philips Cardiac Workstation 7000 recall (2026-07-01), the Covidien Mon-a-Therm Temperature Probe Recall (2026-04-23), and the Philips EPIQ Elite Ultrasound System Recall (2026-08-07).

How to tell if you have this one

Check for the UDI (01)00884838073524 or REF 989605431171 on the transducer. These transducers work with Philips EPIQ 7, Affiniti 70, EPIQ 5, and EPIQ CVxi ultrasound systems.

  • Brand
    Philips
  • Product
    Philips Ultrasound Inc recalls Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5
  • Category
    Medical device
FDA recall event · 4 products

All 4 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    Worldwide distribution - US Nationwide distribution.
Recall numberProduct, codes & lots
Z-1841-2020Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies; All lots.
Z-1842-2020Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies; All lots.
Z-1843-2020Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies; All lots.
Z-1844-2020Transducer S8-3t UDI (01)00884838067523 REF 989605420183 Can be used with: iE33 Utrasound System - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies; All lots.
Based on the official FDA remedy

What to do — step by step

  1. Check Transducer Details

    Look for the UDI (01)00884838073524 or REF 989605431171 on the transducer.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 5.2 years after the firm started it.

1. Firm started the recall: Aug 30, 201912. FDA classified it Class II: May 1, 202023. Published in FDA’s Enforcement Report: May 13, 202034. FDA terminated the recall: Nov 27, 2024420192024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1843-2020 (August 30, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1843-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA