GE Healthcare CardioLab/ComboLab Systems Recall: Risk of Electrical Shock from Patient

GE Healthcare recalls CardioLab/ComboLab Recording Systems with specific model numbers and lot codes due to potential failure in patient leakage current test, which could cause electrical shock if connected to other medical devices.

🇺🇸 FDA recall #Z-0257-2020Medical deviceGE Healthcare ›
Take action

Contact the company

Recalling firmGE Healthcare, LLC3000 N Grandview Blvd · Waukesha, WI 53188-1615
✎ GovInfoNow plain-language explainer

GE Healthcare is recalling certain CardioLab and ComboLab Recording Systems because they may fail to properly test patient leakage current. If these devices connect with other medical devices like ablation systems or ECG monitors, they could allow electrical shocks to patients. Read more at govinfonow.com/recall/67013-ge-healthcare-cardiolab-combolab-systems-recall-risk-of-electrical-shock.

Source Summary by GovInfoNow · govinfonow.com/recall/67013-ge-healthcare-cardiolab-combolab-systems-recall-risk-of-electrical-shock · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0257-2020

Cite this page

“GE Healthcare CardioLab/ComboLab Systems Recall: Risk of Electrical Shock from Patient.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/67013-ge-healthcare-cardiolab-combolab-systems-recall-risk-of-electrical-shock. Based on FDA recall #Z-0257-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0257-2020%22&limit=1).

The FDA announced on 2019-09-16 that GE Healthcare LLC is recalling CardioLab/ComboLab Recording Systems due to a potential patient electrical shock hazard. Owners should contact the company for remedy details.

How to tell if you have it

  • Brand: GE Healthcare
  • Product: CardioLab/ComboLab Recording Systems
  • Model Number: 2088700-002
  • Lot Numbers (Model 2088700-002): RXJ19122006TA, RXJ19122002TA, RXJ19122010TA, RXJ19122001TA, RXJ19122005TA, RXJ19122007TA
  • Model Number: 2088700-004
  • Lot Numbers (Model 2088700-004): RXJ19204004TA, RXJ19154006TA, RXJ19224011TA, RXJ19224017TA, RXJ19224014TA, RXJ19154009TA, RXJ19154011TA, RXJ19154010TA, RXJ19154007TA, RXJ19194009TA, RXJ19154003TA, RXJ19224002TA, RXJ19224016TA, RXJ19204003TA, RXJ19154001TA, RXJ19194003TA, RXJ19224013TA, RXJ19194002TA, RXJ19194011TA, RXJ19204007TA, RXJ19154005TA, RXJ19204006TA, RXJ19194007TA, RXJ19194004TA, RXJ19154008TA, RXJ19204002TA, RXJ19224001TA, RXJ19194001TA, RXJ19224012TA, RXJ19194008TA, RXJ19194005TA, RXJ19154004TA, RXJ19194006TA, RXJ19224009TA, RXJ19204005TA, RXJ19194010TA, RXJ19154002TA

The hazard

The recall addresses the potential for failure of the patient leakage current test. If another device with electrical connectivity fails, the CLab II Plus Amplifier may not prevent an electric current from completing a circuit, which could cause an electrical shock to a patient.

Context

GovInfoNow's records show GE Healthcare LLC has 97 other recalls on record since 2017-10-04. Other recent recalls of similar products include GE Healthcare SIGNA Premier MRI Scanners May Have Steel Fittings Risking Injury (2026-02-27), GE Centricity Universal Viewer 6.0/7.0/8.0: Inaccurate Distance Measurements (2022-07-15), and GE Healthcare SIGNA Premier MRI Scanner Recall: Risk of Hearing Loss from Acoustic Noise (2022-06-24). Other recent recalls of similar products include GE Medical Systems Allia IGS 3 Pulse X-Ray System Recall: Electrical Shock Risk (2026-04-20), GE Medical Systems Allia IGS Pulse X-ray System Recall: Risk of Electrical Shock During Service (2026-04-20), and GE Medical Systems Allia IGS 7 Pulse X-Ray System Recall: Electrical Shock Risk for Service Personnel (2026-04-20).

How to tell if you have this one

Check for model numbers 2088700-002 or 2088700-004 with specific lot numbers listed in the recall notice. These devices were sold with the mentioned lot codes between 2019 and 2020.

  • Brand
    GE Healthcare
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Model

    Identify if your CardioLab or ComboLab Recording System has model numbers 2088700-002 or 2088700-004 with the listed lot numbers.

  2. Stop Using Immediately

    If you have the recalled model with a listed lot number, stop using the device until you contact GE Healthcare for further instructions.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 15 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: nationwideILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: named in the noticeMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: named in the noticeNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Nationwide distribution to CA, CT, FL, IN, MD, MI , MN, MO, NC, NE, NJ, OH, PA, TX, WI. International distribution to Canada, France, India, Japan, Korea (Republic Of), Mexico, Spain, Taiwan.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: California, Connecticut, Florida, Indiana, Maryland, Michigan, Minnesota, Missouri, Nebraska, New Jersey, North Carolina, Ohio, Pennsylvania, Texas, Wisconsin.

Recall timeline

FDA terminated this recall 15 months after the firm started it.

1. Firm started the recall: Sep 16, 201912. FDA classified it Class II: Oct 31, 201923. Published in FDA’s Enforcement Report: Nov 6, 201934. FDA terminated the recall: Dec 16, 2020420192020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0257-2020 (September 16, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0257-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA