Skip to content
Lower risk Medical device recall · FDA · Published October 28, 2019

Genius 2 Tympanic Thermometer Recall: Calibration Drift Risk

Cardinal Health recalls Genius 2 Tympanic Thermometers (OEM Item 303062) made after Oct 2016 due to calibration drift causing inaccurate readings. Affected units may show temperatures higher than specified.

🇺🇸 FDA recall #Z-0992-2020Medical deviceCardinal Health ›
Take action

Contact the company

Recalling firmCardinal Health 200, LLC15 Hampshire St Bldg 5 · Mansfield, MA 02048-1113
✎ GovInfoNow plain-language explainer

Cardinal Health is recalling Genius 2 Tympanic Thermometers made after October 2016 because they may drift out of calibration over time, leading to falsely high temperature readings. This could cause unnecessary medical concerns if used for fever checks. Read more at govinfonow.com/recall/66594-genius-2-tympanic-thermometer-recall-calibration-drift-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/66594-genius-2-tympanic-thermometer-recall-calibration-drift-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0992-2020

Cite this page

“Genius 2 Tympanic Thermometer Recall: Calibration Drift Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/66594-genius-2-tympanic-thermometer-recall-calibration-drift-risk. Based on FDA recall #Z-0992-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0992-2020%22&limit=1).

Cardinal Health 200, LLC recalled Genius 2 Tympanic Thermometers due to potential calibration drift exceeding accuracy tolerances, the FDA said on Oct. 28, 2019. The agency announced the Class II recall under number Z-0992-2020.

How to tell if you have it

  • Brand: Cardinal Health
  • Product: Genius 2 Tympanic Thermometer OEM Tympanic
  • Item Code: 303062
  • Manufacture date: After October 1, 2016
  • Serial Numbers: Greater than or equal to 16597907

The hazard

The FDA stated that the calibration frequency in the operating manual may not keep thermometers within the stated accuracy range. Measurement readings drift upwards over time, potentially exceeding the upper tolerance of +0.2°C for Genius 2 models.

Context

GovInfoNow's records show Cardinal Health 200, LLC has 59 other recalls on record since June 27, 2017. Other recent recalls include Presource Packs with Microstream Components (Aug. 19, 2026), Presource Convenience Kits (Aug. 10, 2026), and Presource Packs Containing Stryker tube sets (Aug. 18, 2026). Similar product recalls include the Covidien Mon-a-Therm Temperature Probe (April 23, 2026) and the Boston Scientific LUX-DX II Arrythmia Detector (March 30, 2026).

How to tell if you have this one

Check for the model name 'Genius 2 Tympanic Thermometer' with OEM Item Code 303062. The product was made after October 1, 2016, and has a serial number greater than or equal to 16597907.

  • Brand
    Cardinal Health
  • Category
    Medical device
FDA recall event · 5 products

All 5 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Martinique Mayotte Mexico Netherlands New Caledonia New Zealand Norway Panama Philippines Poland Portugal Puerto Rico Qatar Reunion Romania San Marino Saudi Arabia Singapore Slovakia Slovenia South Africa South Korea Spain Swaziland Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
Recall numberProduct, codes & lotsQuantity
Z-0990-2020Genius 2 Tympanic Thermometer (discontinued) Item Code: 303000; All product manufactured after October 1, 2016; Serial Numbers greater than or equal to 1659808730142 US; OUS 113841
Z-0991-2020Genius 3 Tympanic Thermometer- electric thermometer Item Code: 303013; All product manufactured after December 4, 2017; Serial Numbers greater than or equal to N177001017425 US; OUS 214278
Z-0992-2020Genius 2 Tympanic Thermometer OEM Tympanic Item Code:303062; All product manufactured after October 1, 2016; Serial Numbers greater than or equal 165979073619 OUS
Z-0993-2020Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063; All product manufactured after December 10, 2017; Serial Numbers greater than or equal to N17606406in total 37567 units US; 335573 OUS
Z-0994-2020Genius 2 Tympanic Thermometer Private Label Item Code: 3069; All product manufactured after October 1, 2016; Serial Numbers greater than N16601357in total 37567 units US; 335573 OUS
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Number

    Look for a serial number starting with 16597907 or higher on the thermometer.

  2. Verify Model

    Confirm the product is the Genius 2 Tympanic Thermometer with OEM Item Code 303062.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 2.6 years after the firm started it.

1. Firm started the recall: Oct 28, 201912. FDA classified it Class II: Feb 4, 202023. Published in FDA’s Enforcement Report: Feb 12, 202034. FDA terminated the recall: May 17, 2022420192022
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Do I need to stop using this thermometer?

No action is required for unaffected units. Affected units may continue use but readings could be inaccurate.

Similar recalls

Other Cardinal Health recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0992-2020 (October 28, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0992-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA