RaySearch recalls specific versions of RayStation treatment planning software due to a bug that may set beam dose to default values, requiring recalibration before treatment approval.
Editing a static arc beam may unexpectedly set the MU of the beam to its initial default value of 200 MU. If the bug is triggered, the dose is invalidated and needs to be recalculated before the plan can be approved or exported.
RaySearch Laboratories is recalling specific versions of their RayStation software treatment planning system. A bug can cause the software to reset dose calculations to default values, meaning treatments may be unsafe if not corrected before approval.
Editing a static arc beam in the software can unexpectedly reset the beam's dose calculation to a default value of 200 MU. This invalidates the dose calculation, requiring a new calculation before treatment plans can be approved or exported. If not corrected, treatments could be unsafe.
Clinically trained radiation therapy staff using RaySearch RayStation software versions 4.0 through 9.0 are most at risk. Affected users must ensure dose calculations are recalculated after any edit to the arc beam.
Check your RayStation software version numbers listed in the recall notice: 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, or 9.0. The software is used for radiation therapy treatment planning by trained medical staff.
Verify your RayStation software version matches the recalled versions listed in the notice.
If you edited a static arc beam, recalculate the dose before approving or exporting the treatment plan.
A bug can cause the software to reset dose calculations to default values after editing a static arc beam, requiring recalibration before treatment approval.
Software versions 4.0 through 9.0, including specific build numbers like 4.0.0.14, 4.5.0.19, and 9.0.0.113.
No, but you must recalculate dose after editing a static arc beam to ensure treatment safety.
We’ve drafted a message you can send RaySearch to request your refund or repair — edit it as you like.
Subject: Recall Z-0736-2020 — RaySearch RaySearch Hello, I own a RaySearch that is covered by recall Z-0736-2020 from RaySearch. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.