The FDA announced on 2019-11-18 that RaySearch Laboratories AB is recalling RayStation stand-alone software treatment planning systems, model numbers 4.0 through 6.3, due to a hazard where editing a static arc beam may unexpectedly reset the MU to 200, invalidating the dose. This recall affects 956 devices distributed nationwide.
How to tell if you have it
- Brand: RaySearch
- Product: RayStation stand-alone software treatment planning system
- Model Numbers: 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3
- Software Versions: 4.0.0.14, 4.0.1.4, 4.0.2.9, 4.0.3.4, 4.5.0.19, 4.5.1.14, 4.5.2.7, 4.7.0.15, 4.7.1.10, 4.7.2.5, 4.7.3.13, 4.7.4.4, 4.7.5.4, 4.9.0.42, 5.0.0.37, 5.0.1.11, 5.0.2.35, 6.0.0.24, 6.1.0.26, 6.1.1.2, 6.2.0.7, 6.3.0.6
- Quantity: 956 Devices/Licences
- Distribution: US Nationwide in AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, VT, WA, and WI
The hazard
According to the FDA, editing a static arc beam may unexpectedly set the MU of the beam to its initial default value of 200 MU. If this bug is triggered, the dose is invalidated and needs to be recalculated before the plan can be approved or exported.
Context
GovInfoNow's records show RaySearch Laboratories AB has had 36 other recalls on record since 2017-07-17. Other recent recalls of similar products include RayCare 2:024A SP1 Oncology Software Data Loss Risk (2026-05-20), RayStation 2024B Software Recall: Radiation Dose Calculation Errors (2025-11-28), and RayStation Software Version 8.1.0.47: Unique DICOM IDs Risk (2025-08-08). Other recent recalls of similar products include Elekta Leksell GammaPlan 11.1 Software Recall: Risk of Incorrect Patient Irradiation (2025-12-12), Boston Scientific TheraSphere 360 Y-90 Software Anomaly Recall (2026-06-19), and Elekta Leksell Vantage Arc System Recall: Risk of shifting coordinates during procedures (2026-03-19).