Skip to content
Use caution Medical device recall · FDA · Published November 25, 2019

Vyaire Medical Bellavista 1000 Ventilator Recall: Intermittent Alarm Failures

Vyaire Medical recalls Bellavista 1000 ventilators with catalog number 301.100.030 due to intermittent failures including lack of high-priority alarms and non-responsive touch screens. Affected units have specific serial numbers listed in the recall.

🇺🇸 FDA recall #Z-1439-2020Medical deviceVyaire Medical ›
Take action

Contact the company

Recalling firmVyaire Medical26125 N Riverwoods Blvd · Mettawa, IL 60045-3420
✎ GovInfoNow plain-language explainer

What happened, in plain English

Vyaire Medical is recalling Bellavista 1000 ventilators with catalog number 301.100.030 because they may experience intermittent failures like missing high-priority alarms or unresponsive touch screens. This could lead to unsafe patient ventilation if the ventilator fails to alert healthcare providers during critical situations. Read more at govinfonow.com/recall/66429-vyaire-medical-bellavista-1000-ventilator-recall-intermittent-alarm-failures.

Source Summary by GovInfoNow · govinfonow.com/recall/66429-vyaire-medical-bellavista-1000-ventilator-recall-intermittent-alarm-failures · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1439-2020

Cite this page

“Vyaire Medical Bellavista 1000 Ventilator Recall: Intermittent Alarm Failures.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/66429-vyaire-medical-bellavista-1000-ventilator-recall-intermittent-alarm-failures. Based on FDA recall #Z-1439-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1439-2020%22&limit=1).

Why it’s dangerous

The ventilator may stop producing high-priority alarms or become unresponsive to touch, meaning healthcare providers might not be alerted to critical issues like airway obstructions or equipment failures. This could lead to unsafe patient ventilation if the ventilator fails to alert staff during emergencies.

Who’s affected

Healthcare providers using the Bellavista 1000 ventilator for patient care are most at risk. Patients receiving ventilation support through this device could be affected if the ventilator malfunctions.

How to tell if you have this one

Check for the catalog number 301.100.030 on the ventilator. Affected units have specific serial numbers listed in the recall, including MB170104 through MB230298.

  • Brand
    Vyaire Medical
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check for Recalled Serial Numbers

    Review the serial numbers listed in the recall to determine if your ventilator is affected.

Where it went and when

Where it was distributed

Distributed nationwide; the notice names 11 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: named in the noticeILWisconsin: nationwideWIMichigan: named in the noticeMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: nationwideOHPennsylvania: named in the noticePANew Jersey: nationwideNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: named in the noticeCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: named in the noticeMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFLPuerto Rico: named in the noticePR

FDA distribution: “US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, California, Colorado, Florida, Georgia, Illinois, Maryland, Michigan, Pennsylvania, Puerto Rico, Texas, Washington.

Recall timeline

FDA terminated this recall 3.7 years after the firm started it.

1. Firm started the recall: Nov 25, 201912. FDA classified it Class II: Mar 4, 202023. Published in FDA’s Enforcement Report: Mar 11, 202034. FDA terminated the recall: Aug 8, 2023420192023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

What should I do if I have this ventilator?

Check the serial number against the list of affected units provided in the recall. If it matches, contact Vyaire Medical for further instructions.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1439-2020 (November 25, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1439-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA