The FDA announced on 2019-12-12 that Arrow International Inc. voluntarily recalled its Arrow Arterial Catheterization Kit due to a labeling error displaying an incorrect Magnetic Resonance symbol, according to the FDA.
How to tell if you have it
- Brand: Arrow
- Product: Arrow Arterial Catherization Kit-20 Ga. x 4.45 cm Radiopaque Polyurethane over 22 Ga. TW Introducer Needle with Integral .018" (0.46 mm) dia. Spring-Wire Guide
- Product Code: ASK-04020-PMC
- Lot/Batch Number: PMC 23F19H0224
- Quantity: 80 eaches
The hazard
The lidstock label incorrectly displays a MR (Magnetic Resonance) Conditional symbol instead of a MR Unsafe symbol in reference to a component of the kit, according to the FDA.
Context
GovInfoNow's records show ARROW INTERNATIONAL Inc. has 52 other recalls on record since 2017-06-01. Other recent recalls include Arrow Arterial Kits and Trays (2024-07-26), the Arrow UltraFlex Intra-Aortic Balloon Catheter Kit (2024-04-29), and the ARROW QuickFlash Radial Artery Catheter (2024-02-12). Additional similar product recalls include the Arrow Radial Artery Catheter Kit (2026-03-11), Arrow PI Midline 2L Catheter (2026-04-21), and Arrow Arterial Line Kits (2026-04-21).