Skip to content
Lower risk Medical device recall · FDA · Published January 10, 2020

Tosoh AIA-360 Analyzer Software Issue: Display Freezes During Use

Tosoh Bioscience recalls AIA-360 Automated Immunoassay Analyzers with software issue causing display freezes when touched during command execution, leading to aborted runs and unretrievable results.

🇺🇸 FDA recall #Z-1420-2020Medical deviceTosoh Bioscience ›
Take action

Contact the company

Recalling firmTosoh Bioscience Inc3600 Gantz Rd · Grove City, OH 43123-1895
✎ GovInfoNow plain-language explainer

Tosoh Bioscience is recalling AIA-360 Automated Immunoassay Analyzers due to a software issue that causes the display to freeze when touched while the instrument is processing commands. This results in test runs being aborted and results being unrecoverable. The issue does not pose a physical safety risk but affects test accuracy. Read more at govinfonow.com/recall/66112-tosoh-aia-360-analyzer-software-issue-display-freezes-during-use.

Source Summary by GovInfoNow · govinfonow.com/recall/66112-tosoh-aia-360-analyzer-software-issue-display-freezes-during-use · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1420-2020

Cite this page

“Tosoh AIA-360 Analyzer Software Issue: Display Freezes During Use.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/66112-tosoh-aia-360-analyzer-software-issue-display-freezes-during-use. Based on FDA recall #Z-1420-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1420-2020%22&limit=1).

The FDA announced on Jan. 10, 2020, that Tosoh Bioscience Inc is recalling the AIA-360 Automated Immunoassay Analyzer due to a display screen software issue. The hazard causes the instrument to stop and aborts runs, making results unretrievable.

How to tell if you have it

  • Brand: Tosoh Bioscience
  • Product: AIA-360 Automated Immunoassay Analyzer
  • Product Code: 019945
  • Serial Numbers: 28469102, 28629205, 28469202, 28629305, 28469302, 28629405, 28469402, 28629505, 28469502, 28629605, 28469602, 28629705, 28469702, 28709907, 28469802, 28719007, 28469902, 28719107, 28479002, 28719207, 28519303, 28719307, 28519403, 28719407, 28519503, 28719507, 28519603, 28719607, 28519703, 28719707, 28519803, 28719807, 28519903, 28719907, 28529003, 28729007, 28529103, 28729108, 28529203, 28729208, 28529903, 28729308, 28539003, 28729408, 28539103, 28729508, 28539203, 28729608, 28539303, 28759908, 28539403, 28769008, 28539503, 28769108, 28539603, 28769208, 28539703, 28769308, 28539803, 28769408
  • Quantity: 12 units
  • Distribution: AZ, NE, NJ, NY, WA, Cayman Islands, Costa Rica, Uruguay

The hazard

A display screen software issue causes the analyzer's display to freeze when touched simultaneously with a firmware command. This stops the instrument, aborts the run, and prevents retrieval of results.

Context

GovInfoNow's records show Tosoh Bioscience Inc has 19 other recalls on record since June 11, 2017. Other recent recalls of similar products include Tosoh ST AIA-Pack Testosterone Recall (June 12, 2026) and Tosoh AIA-Pack BHCG Calibrator Recall (Aug. 7, 2020). AU480 Chemistry Analyzer May Show Error Message Causing Delay in Patient Results was reported on Aug. 3, 2026.

How to tell if you have this one

Check for the AIA-360 Automated Immunoassay Analyzer with Product Code 019945. The device was sold with specific serial numbers listed in the recall notice.

  • Brand
    Tosoh Bioscience
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Status

    Verify if the display freezes when touched during command execution.

Where it went and when

Where it was distributed

Distributed in 5 states, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: yesWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: yesNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: yesNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: yesNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: yesAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: not listedFL

FDA distribution: “Distributed to AZ, NE, NJ, NY, and WA. Foreign distribution to Cayman Islands, Costa Rica, and Uruguay” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Arizona, Nebraska, New Jersey, New York, Washington.

Recall timeline

FDA terminated this recall 8 months after the firm started it.

1. Firm started the recall: Jan 10, 202012. FDA classified it Class II: Mar 2, 202023. Published in FDA’s Enforcement Report: Mar 11, 202034. FDA terminated the recall: Sep 10, 202042020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1420-2020 (January 10, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1420-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA