The FDA announced on 2020-02-18 that Microbiologics Inc is recalling QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set due to a risk of incorrect susceptible test results. The agency classified the recall as Class II.
How to tell if you have it
- Brand: Microbiologics
- Product: QC Sets and Panels, KWIK-STIK AST-GP (4 Strains) QC Set
- Catalog Number: 0959P
- Catalog Number: 5112P
- Lot Numbers: 5112-01 (Expiration Date 07/31/2020), 5112-02 (Expiration Date 10/31/2020)
- UDI: 70845357026551
The hazard
GovInfoNow's records show the recalled kits may return incorrect susceptible results. The high level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Context
GovInfoNow's records show Microbiologics Inc has 35 other recalls on record since 2017-06-06. Other recent recalls of similar products include Microbiologics KWIK-STIK 6-Pack Recall (2025-12-02), Microbiologics KWIK-STIK Quality Control Kit Recall (2025-11-13), and Microbiologics LYFO-DISK Recall (2025-06-27).