Skip to content
Use caution Medical device recall · FDA · Published February 20, 2020

Radiometer TCM4 Base Unit Recall: Display Flickering May Delay Patient Monitoring

Radiometer recalls TCM4 Base units with specific serial numbers due to display flickering and monitor unresponsiveness, which could delay patient monitoring and compromise product performance.

🇺🇸 FDA recall #Z-1755-2020Medical deviceRadiometer ›
Take action

Contact the company

Recalling firmRadiometer Medical ApSAkandevej 21 · Bronshoj, Denmark
✎ GovInfoNow plain-language explainer

Radiometer is recalling specific TCM4 Base units that may flicker or become unresponsive, potentially delaying patient monitoring and causing unsafe monitoring conditions. This affects medical devices used for pediatric and adult patient monitoring. Read more at govinfonow.com/recall/65803-radiometer-tcm4-base-unit-recall-display-flickering-may-delay-patient.

Source Summary by GovInfoNow · govinfonow.com/recall/65803-radiometer-tcm4-base-unit-recall-display-flickering-may-delay-patient · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1755-2020

Cite this page

“Radiometer TCM4 Base Unit Recall: Display Flickering May Delay Patient Monitoring.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/65803-radiometer-tcm4-base-unit-recall-display-flickering-may-delay-patient. Based on FDA recall #Z-1755-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1755-2020%22&limit=1).

The FDA announced on Feb. 20, 2020, that Radiometer Medical ApS is recalling TCM4 Base units due to display flickering or unresponsiveness that may delay patient monitoring. Owners should contact the company for a remedy.

How to tell if you have it

  • TCM4 Base unit, Ref 391-880
  • UDI: (01)05700693918808
  • UPC: 05700693918808
  • Used with monitor modules TCM4 (902-778), TCM40 (902-778, 902-946, or 902-862), TCM400 (902-564), TOSCA (903-044) and CombiM (903-111)
  • Affected Serial Numbers: 391-880R0471N017, 391-880R0471N030, 391-880R0473N006, 391-880R0473N016, 391-880R0471N021, 391-880R0472N010, 391-880R0473N007, 391-880R0471N022, 391-880R0472N011, 391-880R0473N008, 391-880R0471N023, 391-880R0472N014, 391-880R0473N009, 391-880R0471N024, 391-880R0472N016, 391-880R0473N010, 391-880R0471N025, 391-880R0472N020, 391-880R0473N011, 391-880R0471N026, 391-880R0473N001, 391-880R0473N020, 391-880R0471N027, 391-880R0473N002, 391-880R0473N013, 391-880R0471N028, 391-880R0473N003, 391-880R0473N014, 391-880R0471N029, 391-880R0473N005, 391-880R0473N015
  • Sold worldwide including FL, KS, NY, ND, IN, TX, CA, VA, MA, MN, WI, CT, WA, China, India, Japan, and South Korea

The hazard

Reports indicate the transcutaneous monitoring system display may flicker or become non-responsive and beep. This issue may result in delayed patient monitoring and potentially compromise product performance.

Context

GovInfoNow's records show Radiometer Medical ApS has 16 other recalls on record since Feb. 27, 2018. Other recent recalls of similar products include Philips IntelliVue MX40: Patient Monitor May Not Alarm (Oct. 31, 2025), Philips IntelliVue Multi-Measurement Module X3 Recall: Monitor May Not Alarm (Oct. 31, 2025) and Philips Telemetry Monitor 5500 1.4 GHz Recall: Risk of Device Reset Causing Connection Failure (June 5, 2026).

How to tell if you have this one

Check if your TCM4 Base unit has the serial numbers listed in the recall notice (e.g., 391-880R0471N017). The device is used with specific monitor modules including TCM4, TCM40, TCM400, TOSCA, and CombiM models.

  • Brand
    Radiometer
  • Product
    TCM4 Base unit, Ref 391-880, UDI:(01)05700693918808, used with the monitor modules TCM4 (902-778), TCM40 (902-778
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Serial Number

    Identify your TCM4 Base unit's serial number to see if it matches the affected list provided in the recall notice.

  2. Contact Radiometer for Guidance

    Reach out to Radiometer Medical ApS directly for further instructions on whether your device needs repair or replacement.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 13 states.

Alaska: nationwideAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: named in the noticeNDMinnesota: named in the noticeMNIllinois: nationwideILWisconsin: named in the noticeWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: named in the noticeINOhio: nationwideOHPennsylvania: nationwidePANew Jersey: nationwideNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: nationwideMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: named in the noticeVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “Worldwide distribution - US Nationwide distribution including the states of FL, KS, NY, ND, IN, TX, CA , VA, MA, MN, WI, IN, CT, WA and the countries of China, India, Japan, south Korea.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: California, Connecticut, Florida, Indiana, Kansas, Massachusetts, Minnesota, New York, North Dakota, Texas, Virginia, Washington, Wisconsin.

Recall timeline

FDA terminated this recall 2.3 years after the firm started it.

1. Firm started the recall: Feb 20, 202012. FDA classified it Class II: Apr 22, 202023. Published in FDA’s Enforcement Report: Apr 29, 202034. FDA terminated the recall: Jun 24, 2022420202022
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1755-2020 (February 20, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1755-2020 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA