Abbott recalled the Merlin PCS 3650 Software Upgrade Kit (REF 3330) on Feb. 28, 2020, because a software anomaly may cause the device to retain a previously programmed energy level without alerting the user, according to the FDA.
How to tell if you have it
- Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical
- UPC 05414734500661
- UPC 05414734509725
The hazard
Internal testing identified software anomaly scenarios where selecting a non-applicable shock energy value causes the device to continue using the previously programmed energy level. The user will not be alerted that the selected value was not retained.
Context
GovInfoNow's records show Abbott has 12 other recalls on record since April 5, 2018. Other recent recalls of similar products include Abbott TactiFlex Sensor Catheter Recall (Sept. 10, 2025) and Abbott Amplatzer TorqVue LP Delivery System Recall (March 17, 2025). Medtronic issued a software issue recall for the Restore CP App Model A71100 on Nov. 20, 2025, and Fresenius Kabi recalled the Ivenix Infusion System on May 6, 2026.