Canon recalls Infinix-8000C catheterization tables with CAT-850B or CAT-860B models due to loose bolts causing table tilting. Affected units have specific serial numbers listed in the recall notice.
Bolts holding the longitudinal movement of catheterization tables may be loose, which may lead to table surface/top tilting.
Canon is recalling certain Infinix-8000C catheterization tables (CAT-850B or CAT-860B models) because bolts holding the table may be loose, potentially causing the table to tilt. This could lead to injury if a patient or medical staff falls during procedures.
Loose bolts in the table's longitudinal movement mechanism can cause the table surface to tilt unexpectedly. This tilting could lead to falls or injuries for patients or medical staff during procedures if the table is not properly secured.
Healthcare professionals and patients using the recalled catheterization tables for medical procedures are most at risk. The specific units affected have unique serial numbers listed in the recall notice.
Check if your Infinix-8000C catheterization table has a serial number starting with B8B, A9A, A3A, AEB, A1B, A9C, A3E, B5B, A9B, A5C, B3B, A1D, B1B, A5C, A9C, A1E, A1D, A1E, A5E, AEA.
Identify the serial number on your catheterization table to confirm if it is listed in the recall notice.
If your table has a recalled serial number, stop using it immediately to prevent tilting and potential injury.
The recall affects Infinix-8000C catheterization tables with CAT-850B or CAT-860B models.
Loose bolts in the table's longitudinal movement mechanism can cause the table surface to tilt unexpectedly, potentially leading to falls or injuries.
Check the serial number on your table; affected units have specific serial numbers starting with B8B, A9A, A3A, AEB, A1B, A9C, A3E, B5B, A9B, A5C, B3B, A1D, B1B, A5C, A9C, A1E, A1D, A1E, A5E, or AEA.
We’ve drafted a message you can send Canon Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-2326-2020 — Canon Medical Canon Medical Hello, I own a Canon Medical that is covered by recall Z-2326-2020 from Canon Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.