Canon recalls specific catheterization tables (INFX-8000C models) due to loose bolts that may cause the table to tilt. Affected units include certain serial numbers and model combinations.
System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Canon is recalling specific catheterization tables that may tilt because bolts holding the vertical column and table top can become loose. This could lead to instability during medical procedures, potentially causing injury if the table tips.
The bolts that secure the vertical column and the table top tilting drive unit can become loose over time. If the table tilts, it could cause the patient to fall or the table to shift unexpectedly during medical procedures, leading to potential injury.
Healthcare professionals using the Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000C catheterization tables with the specified serial numbers and model combinations are most at risk. Patients undergoing procedures on these tables could be affected if the table tilts unexpectedly.
Check for the model numbers INFX-8000C/B6, INFX-8000C/B4, or other listed variants. The table serial numbers listed in the recall include codes like B1C2072072, B1C2072070, and others starting with 99A or B1B.
Look for the table serial number on the product label, which may start with B1C, B1B, 99A, or 99B.
Reach out to Canon Medical System, USA, INC. for assistance with the recall and potential remedies.
The bolts that fasten the vertical column and table top tilting drive unit may become loose, causing the table to tilt during use. This could lead to instability and potential injury.
The affected models include Canon INTERVENTIONAL ANGIOGRAPHY SYSTEM MODEL INFX-8000C with specific variants like INFX-8000C/B6, INFX-8000C/B4, and others listed in the recall.
Check the table serial number on the product label. Affected units have serial numbers starting with B1C, B1B, 99A, or 99B as listed in the recall notice.
We’ve drafted a message you can send Canon Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1270-2021 — Canon Medical Canon Medical Hello, I own a Canon Medical that is covered by recall Z-1270-2021 from Canon Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.