Canon recalls specific catheterization tables (INFX-8000F model) due to loose bolts that could cause table tilting. Affected units have unique serial numbers; contact Canon for remedy.
System table may tilt because the bolts that fasten the vertical column and the table top tilting drive unit may become loose.
Canon is recalling certain catheterization tables that may tilt because bolts holding the table top and vertical column can become loose. This could lead to injury if the table tips during medical procedures.
The bolts that secure the vertical column and table top tilting mechanism may loosen over time. If the table tilts while in use during medical procedures, it could cause injury to the patient or healthcare worker.
Healthcare professionals using Canon's INFX-8000F catheterization tables with specific serial numbers are most at risk. Patients undergoing procedures on these tables could be affected if the table tilts unexpectedly.
Check for the model INFX-8000F with serial numbers listed in the recall notice. The table has a specific serial number format starting with letters like '99E' or '99F' followed by numbers.
Look for the serial number on the table, which starts with letters like '99E' or '99F' followed by numbers.
Reach out to Canon Medical System, USA, INC. to confirm if your table is affected and get instructions for repair or replacement.
The bolts that fasten the vertical column and table top tilting drive unit may become loose, causing the table to tilt during medical procedures.
Check the serial number on the table, which starts with letters like '99E' or '99F' followed by numbers. Affected models are INFX-8000F with specific serial numbers listed in the recall notice.
Contact Canon Medical System, USA, INC. to confirm if your table is affected and get instructions for repair or replacement.
We’ve drafted a message you can send Canon Medical to request your refund or repair — edit it as you like.
Subject: Recall Z-1271-2021 — Canon Medical Canon Medical Hello, I own a Canon Medical that is covered by recall Z-1271-2021 from Canon Medical. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.