The FDA announced on March 4, 2020, that RaySearch Laboratories AB is recalling RayStation standalone software treatment planning systems due to potential radiation dose errors. The agency stated the hazard involves restricted maximum field openings for Siemens Virtual Wedge fields.
How to tell if you have it
- Brand: RaySearch
- Product: RayStation standalone software treatment planning system
- Model Numbers: 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 7.0, 8.0, 8.1
- Units affected: 1389 licenses
The hazard
According to the FDA, the physical limitations of the treatment machine cause the maximum field opening for Siemens Virtual Wedge fields to be more restricted than recommended. Delivering a larger field opening with a virtual wedge results in a high dose plateau region at the toe side of the field beyond the recommended field size instead of a wedged field shape.
Context
GovInfoNow's records show RaySearch Laboratories AB has issued 36 other recalls since July 17, 2017. Recent similar recalls include the RayCare 2:024A SP4 Oncology Software Data Loss Risk (May 20, 2026), the RayStation 10B Software Radiation Dose Errors Recall (Nov. 28, 2025), and the RayStation Software Version 13.0.0.1547 Unique DICOM IDs Risk Recall (Aug. 8, 2025). Other recent recalls of similar products include the Hitachi Proton Beam Therapy System Positioning Software Anomaly Risk (Feb. 6, 2026) and the Elekta Leksell GammaPlan 11.1 Software Incorrect Patient Irradiation Recall (Dec. 12, 2025).