Respironics California, LLC recalled Philips V60 Ventilators with Power Management PCBA part number 1055906 due to a solder connection failure that may cause unexpected shutdowns or loss of power, the FDA said on March 5, 2020.
How to tell if you have it
- Brand: Respironics
- Product: Philips V60 Ventilators with Power Management PCBA part number 1055906
- Serial Numbers: 100002908 to 100027093; 100027096 to 100085389; 100085399 to 100109746; 100109868 to 201003390; 201003392 to 201010952
- Quantity: 23,581 units
The hazard
A solder connection failure on the Power Management printed circuit board assembly may cause the blower to lose power and trigger a High Priority Check Vent alarm, or cause an intermittent connection that disrupts operation and triggers an unexpected shutdown without alarm. These failures may lead to hypercarbia or hypoxemia.
Context
GovInfoNow's records show Respironics California, LLC has issued seven other recalls since September 11, 2019. Other recent recalls of similar products include the Philips Respironics V60 Ventilator Recall regarding an internal circuit issue (February 28, 2022), the Philips Respironics V60 Ventilator Recall concerning expired adhesive (January 24, 2022), and the Philips Respironics V60 Plus Ventilator Recall for software versions 3.00 and 3.10 (June 18, 2021). Other recent recalls of similar products include the Baxter Blue Ventilator Adapter Module Recall for oxygen desaturation risk (February 18, 2026), the Philips Respironics Trilogy EV300 Ventilator Software Update Risk (March 2, 2026), and the Ventec V+Pro Ventilator Recall for oxygen leak risk (March 25, 2026).