Lower risk Medical device recall · FDA · Published May 8, 2020

uCT 530 CT Scanner Recall: False Marking and Radiation Risks

Shanghai United Imaging Healthcare recalls uCT 530 CT scanners with software version R001.11.0.729741-Re-1100 due to potential false channel marking and radiation exposure risks from rescan needs.

Take action

Contact the company

Recalling firmShanghai United Imaging Healthcare Co., Ltd.No. 2258 Chengbei Rd, Jiading Ind. · Shanghai, China
✎ GovInfoNow plain-language explainer

Shanghai United Imaging Healthcare is recalling uCT 530 CT scanners with specific software versions. These units may cause false channel markings leading to rescans, which could increase radiation exposure and contrast medium use. Read more at govinfonow.com/recall/65201-uct-530-ct-scanner-recall-false-marking-and-radiation-risks.

Source Summary by GovInfoNow · govinfonow.com/recall/65201-uct-530-ct-scanner-recall-false-marking-and-radiation-risks · Updated August 19, 2026 · FDA enforcement record: FDA #Z-2472-2020

Cite this page

“uCT 530 CT Scanner Recall: False Marking and Radiation Risks.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/65201-uct-530-ct-scanner-recall-false-marking-and-radiation-risks. Based on FDA recall #Z-2472-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2472-2020%22&limit=1).

Shanghai United Imaging Healthcare Co., Ltd. voluntarily recalled four uCT 530 Computed Tomography X-Ray Systems on May 8, 2020, according to the FDA. The devices may produce false markings or scanning interruptions, potentially requiring patient rescans and additional radiation exposure.

How to tell if you have it

  • uCT 530 Computed Tomography X-Ray System
  • Software version R001.11.0.729741-Re-1100
  • Serial numbers: 300162, 300163, 300174, 300176
  • GTINs: 6971576831012
  • UPC: 06971576831012

The hazard

The recall cites two issues: a service function that may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.

Context

GovInfoNow's records show Shanghai United Imaging Healthcare Co., Ltd. has 14 other recalls on record since July 30, 2019. Other recent recalls of similar products include Philips IQon Spectral CT Software 4.7.7.x Recall (Aug. 7, 2026), Philips Brilliance CT 64 Medical Scanner Recall (Aug. 7, 2026), and Philips Ingenuity CT Scanners: Unintended Radiation Risk (Aug. 7, 2026).

How to tell if you have this one

Check for serial numbers 300162, 300163, 300174, or 300176 with software version R001.11.0.729741-Re-1100.

  • Brand
    Shanghai United Imaging Healthcare
  • Product
    uCT 530 Computed Tomography X-Ray System
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version on the device; affected units have version R001.11.0.729741-Re-1100.

Where it went and when

Where it was distributed

Distributed in Texas.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: yesTXFlorida: not listedFL

FDA distribution: “Distributed to Texas.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Texas.

Recall timeline

FDA published this recall 2 months after the firm started it.

1. Firm started the recall: May 8, 202012. FDA classified it Class II: Jun 30, 202023. Published in FDA’s Enforcement Report: Jul 8, 202032020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2472-2020 (May 8, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2472-2020 to see it in context.