The Food and Drug Administration announced on May 18, 2020, that Carefusion 2200 Inc is recalling Mueller Knife Handles due to a hazard involving loose surgical blades. The agency classified the recall as Class II.
How to tell if you have it
- Brand: Carefusion
- Product: Mueller Knife Handle, Hexagonal
- Model Numbers: OP0702 and OP0703
- Lot Codes: C20VAA (4 inch length) and D17VAA (6 inch length)
- Units affected: 358
The hazard
Becton Dickinson and Company is conducting the recall because external threads on some collets contain a machining burr. This defect may prevent the collet from fully seating in the knife handle, inhibiting a complete grip of the inserted blade and possibly resulting in a loose blade.
Context
GovInfoNow's records show Carefusion 2200 Inc has 8 other recalls on record since 2013-12-20. Other recent recalls include a Carefusion Endoscopic Grasper recalled on 2023-07-13 for cracking risk and a Carefusion Vaginal Specula recalled on 2023-03-16 due to label mismatch. Similar product recalls include D.O.R.C. TDC VELOCE cutters recalled on 2026-04-30 for loose knife risks.