RaySearch Laboratories recalls RayStation software versions 8A-9B due to potential treatment data import errors in radiation therapy systems. Affected users should check their software version and contact support for updates.
It is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. In addition, there is an issue found with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.
RaySearch Laboratories is recalling specific versions of their RayStation treatment planning software because importing treatment records into PACS may not correctly transfer data to RayTreat. This could lead to incorrect radiation therapy planning if not addressed.
The software may not properly import treatment records from RayCare PACS into RayTreat, causing data mismatches. Additionally, gantry angle displays in beam lists might be inaccurate, potentially leading to incorrect radiation delivery.
Radiation therapy professionals using RayStation software versions 8A-9B (including all service packs) are affected. Those handling treatment plans for cancer patients are most at risk.
Check your RayStation software version number. Affected versions include 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, and 9.2.0.483. The software was sold with model numbers 8.0, 8.1, 9.0, 9.1, and 9.2.
Review your RayStation software version to see if it matches the affected versions listed in the recall notice.
The recall is due to potential errors in importing treatment records from RayCare PACS into RayTreat and inaccurate display of gantry angles in beam lists.
Check your RayStation software version number against the affected versions listed: 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, and 9.2.0.483.
No action is required if your software version is not listed as affected in the recall notice.
We’ve drafted a message you can send RaySearch Laboratories to request your refund or repair — edit it as you like.
Subject: Recall Z-2426-2020 — RaySearch Laboratories RaySearch Laboratories Hello, I own a RaySearch Laboratories that is covered by recall Z-2426-2020 from RaySearch Laboratories. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.