The FDA announced on May 26, 2020, that RaySearch Laboratories AB is recalling RayStation standalone software treatment planning systems (versions 8A, 8B, 9A, and 9B) due to data import and display errors. According to the FDA, the hazard involves manual treatment records not automatically importing into RayTreat and incorrect Gantry Angle displays.
How to tell if you have it
- Brand: RaySearch Laboratories
- Product: RayStation standalone software treatment planning system
- Models: RayStation 8A, 8B, 9A, and 9B (including all service packs)
- Model numbers: 8.0, 8.1, 9.0, 9.1, 9.2
- Software versions: 8.0.0.61, 8.0.1.10, 8.1.0.47, 8.1.1.8, 8.1.2.5, 9.0.0.113, 9.1.0.933, and 9.2.0.483
- Units affected: 12
The hazard
According to the FDA, it is not obvious that the manual import of a treatment record into RayCare PACS does not automatically import the same data into RayTreat. Additionally, there is an issue with the display of planned and recorded Gantry Angle in the delivered beam list in RayTreat.
Context
GovInfoNow's records show RaySearch Laboratories AB has issued 36 other recalls since July 17, 2017. Recent recalls include the May 20, 2026 RayCare 2:024A SP4 Oncology Software Data Loss Risk and the November 28, 2025 RayStation 10B Software recall for potential radiation dose errors. Other recent recalls of similar products include the December 12, 2025 Elekta Leksell GammaPlan 11.1 Software Recall and the May 21, 2026 Elekta MOSAIQ Oncology System Software Recall.