Skip to content
Use caution Medical device recall · FDA · Published June 19, 2020

Encore Medical EMPOWR Partial Knee Peg Drill Guide Recall: Drilling Risk

Encore Medical recalls EMPOWR Partial Knee Peg Drill Guide model 800-06-007 due to potential drilling issues that may delay surgery. All units are affected.

🇺🇸 FDA recall #Z-0479-2021Medical deviceEncore Medical ›
Take action

Contact the company

Recalling firmEncore Medical, LP9800 Metric Blvd · Austin, TX 78758-5445
✎ GovInfoNow plain-language explainer

Encore Medical is recalling its EMPOWR Partial Knee Peg Drill Guide model 800-06-007 because it can bind during femoral posterior peg drilling, potentially delaying surgery. This risk affects surgeons and medical professionals using the device. Read more at govinfonow.com/recall/64946-encore-medical-empowr-partial-knee-peg-drill-guide-recall-drilling.

Source Summary by GovInfoNow · govinfonow.com/recall/64946-encore-medical-empowr-partial-knee-peg-drill-guide-recall-drilling · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0479-2021

Cite this page

“Encore Medical EMPOWR Partial Knee Peg Drill Guide Recall: Drilling Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/64946-encore-medical-empowr-partial-knee-peg-drill-guide-recall-drilling. Based on FDA recall #Z-0479-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0479-2021%22&limit=1).

Encore Medical, LP recalled the EMPOWR Partial Knee Peg Drill Guide (Model 800-06-007) on June 19, 2020, according to the FDA. The device may bind during surgery, potentially causing delays.

How to tell if you have it

  • Brand: Encore Medical
  • Product: EMPOWR Partial Knee Peg Drill Guide
  • Model: 800-06-007
  • Lot numbers: all lot numbers
  • Quantity: 28 units
  • Distribution: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA

The hazard

The FDA stated that the shaft portion of the drill may bind within the Posterior Peg drill guide during femoral posterior peg preparation. This issue could delay surgery if it recurs.

Context

GovInfoNow's records show Encore Medical, LP has issued 43 other recalls since May 31, 2017. Recent brand recalls include an EMPOWR 3D Knee Ins 11L 14MM Product Mix-Up Recall on Feb. 16, 2026, and an RSP Humeral Socket Insert Recall regarding incorrect labeling on Jan. 15, 2026. Other recent recalls of similar products include a Merit Medical Custom Procedure Kit Recall (Dec. 4, 2025) and Surgify Halo Drill Risks (April 30, 2026).

How to tell if you have this one

Check for the model number 800-06-007 on the EMPOWR Partial Knee Peg Drill Guide. All units produced are affected, with no specific lot numbers mentioned.

  • Brand
    Encore Medical
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.
Recall numberProduct, codes & lots
Z-0479-2021EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.; all lot numbers
Z-0480-2021EMPOWR Partial Knee Peg Drill; Model: 800-06-008.; all lot numbers
Based on the official FDA remedy

What to do — step by step

  1. Check Device Model

    Verify the model number on the drill guide to confirm it is 800-06-007.

  2. Contact Manufacturer

    Reach out to Encore Medical for guidance on next steps if you have the recalled model.

Where it went and when

Where it was distributed

Distributed in 12 states.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: yesWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: yesMNIllinois: yesILWisconsin: not listedWIMichigan: yesMINew York: yesNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: yesINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: yesCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: yesVAMaryland: not listedMDDelaware: not listedDEArizona: yesAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: yesGAHawaii: not listedHITexas: yesTXFlorida: yesFL

FDA distribution: “US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Arizona, California, Florida, Georgia, Illinois, Indiana, Michigan, Minnesota, New York, Texas, Virginia, Washington.

Recall timeline

FDA published this recall 5 months after the firm started it.

1. Firm started the recall: Jun 19, 202012. FDA classified it Class II: Nov 20, 202023. Published in FDA’s Enforcement Report: Dec 2, 202032020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0479-2021 (June 19, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0479-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA