The FDA announced on August 6, 2020, that Carl Zeiss Meditec AG is recalling its VISUREF150 diagnostic instrument due to an electrical shock hazard. The recall involves 132 units distributed worldwide.
How to tell if you have it
- Brand: Carl Zeiss Meditec
- Product: VISUREF150
- Catalog Number: 000000-2227-967
- Serial Numbers: Pending
The hazard
The recall is due to an incorrect DC/DC converter being assembled onto the communication board, which may result in insufficient safety functions and increased current leakage. This defect potentially causes electrical shocks to the user.
Context
GovInfoNow's records show Carl Zeiss Meditec AG has had five other recalls since August 30, 2017. Recent brand recalls include a June 1, 2026, recall of the Zeiss Intrabeam Spherical Applicator for manufacturing defects and a March 16, 2022, recall of the ZEISS miLOOP Lens Fragmentation Device due to high friction risk. Other recent recalls of similar products include a December 15, 2025, recall of the Philips Allura Xper FD10C Medical System for electrical short-circuit risks and an April 20, 2026, recall of the GE Medical Systems Allia IGS 3 Pulse X-Ray System for electrical shock risk.