Use caution Medical device recall · FDA · Published September 1, 2020

Fresenius AquaBplus Water Purification Systems Recall: Electrical Shock Risk

Fresenius recalls AquaBplus and AquaBplus B2 water purification systems for hemodialysis use due to electrical safety risks from loose crimp connections that could cause shock to staff.

🇺🇸 FDA recall #Z-0098-2021Medical deviceFresenius ›
Take action

Contact the company

Recalling firmFresenius Medical Care Holdings, Inc.920 Winter St Bld 950 · Waltham, MA 02451-1521
✎ GovInfoNow plain-language explainer

Fresenius is recalling AquaBplus and AquaBplus B2 water purification systems used in hemodialysis. The recall is because improperly crimped electrical connectors could become loose during movement or maintenance, potentially causing electrical shock to medical staff. Read more at govinfonow.com/recall/64384-fresenius-aquabplus-water-purification-systems-recall-electrical-shock-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/64384-fresenius-aquabplus-water-purification-systems-recall-electrical-shock-risk · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0098-2021

Cite this page

“Fresenius AquaBplus Water Purification Systems Recall: Electrical Shock Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/64384-fresenius-aquabplus-water-purification-systems-recall-electrical-shock-risk. Based on FDA recall #Z-0098-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0098-2021%22&limit=1).

Fresenius Medical Care Holdings, Inc. recalled AquaBplus and AquaBplus B2 Water Purification Systems on Sept. 1, 2020, according to the FDA. The agency announced the Class II recall due to incorrectly crimped connectors that could cause electrical shock.

How to tell if you have it

  • AquaBplus 1500 - Typ USA - 3 x 208V / 60Hz - Fresenius Part Number: 24-1500-1
  • AquaBplus B2 2000 / B1 2500 (Bplus DUO 2000) - 3 x 208V / 60Hz - Fresenius Part Number: 24-2000-0; 9BDS1515 08D51544 OBDS1596 OBPS2864 9BDS1529 OBDS1545 OBPS2799 OBPS2884 9BPS2735 OBDS1551 OBPS2800 OBPS2903 9BPS2767 08D51590 OBPS2810

The hazard

Incorrectly crimped connectors were installed on the main protective earth conductor. If the machine is moved or maintenance is performed, these connections could become loose and adversely affect electrical safety, potentially resulting in electrical shock to Biomed technicians or clinic staff.

Context

GovInfoNow's records show Fresenius Medical Care Holdings, Inc. has 44 other recalls on record since April 14, 2014. Other recent recalls of similar products include Fresenius Bicarby Dialysate Leaks Risk (May 12, 2026), Fresenius 5008X Blood Tubing Set Recall (May 11, 2026), and Fresenius Medical Care Bicarby Dialysate Recall (May 7, 2026). Additional recent recalls include Fresenius Medical Care 5008X Blood Tubing Set Recall (Aug. 10, 2026), Fresenius Medical Care CombiSet Hemodialysis Tubing Set Recall (July 29, 2026), and Fresenius Medical Care 2008T BlueStar Hemodialysis Machine Recall (June 30, 2026).

How to tell if you have this one

Check for the AquaBplus 1500 model (Fresenius Part Number: 24-1500-1) or AquaBplus B2 2000/B1 2500 models (Fresenius Part Numbers: 24-2000-0, 9BDS1515, 08D51544, etc.) sold for hemodialysis use.

  • Brand
    Fresenius
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check for Loose Connections

    Inspect the main protective earth conductor for loose crimp connections that could cause electrical shock during machine movement or maintenance.

Where it went and when

Where it was distributed

Distributed in 5 states.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: yesPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: yesDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: yesNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: yesOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: yesTXFlorida: not listedFL

FDA distribution: “US distribution to DE, NC, OK, PA, and TX.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Delaware, North Carolina, Oklahoma, Pennsylvania, Texas.

Recall timeline

FDA terminated this recall 5 months after the firm started it.

1. Firm started the recall: Sep 1, 202012. FDA classified it Class II: Oct 13, 202023. Published in FDA’s Enforcement Report: Oct 21, 202034. FDA terminated the recall: Jan 20, 2021420202021
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Do I need to stop using the product immediately?

No. The recall requires checking for loose connections but does not mandate immediate discontinuation. Medical staff should follow safety procedures during maintenance.

Similar recalls

Other Fresenius recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0098-2021 (September 1, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0098-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA