The FDA announced on 2020-09-30 that Maquet Cardiovascular Us Sales, Llc is recalling the Blood Monitoring Unit (BMU 40), model no. 701040852, due to non-compliance with IEC standards regarding creepage distance on the printed circuit board assembly.
How to tell if you have it
- Brand: Maquet
- Product: Blood Monitoring Unit (BMU 40)
- Model no.: 701040852
- Serial numbers: 90002022, 90002134, 90002163, 90002187, 90002276, 90002297, 90002023, 90002135, 90002164, 90002188, 90002277, 90002298, 90002024, 90002136, 90002168, 90002191, 90002279, 90002302, 90002029, 90002137, 90002170, 90002194, 90002280, 90002306, 90002047, 90002140, 90002172, 90002195, 90002281, 90002312, 90002112, 90002143, 90002181, 90002196, 90002287, 90002313, 90002123, 90002144, 90002183, 90002210, 90002289, 90002124, 90002160, 90002184, 90002212, 90002293, 90002133, 90002161, 90002185, 90002273, 90002296
- Units affected: 51
The hazard
According to the FDA, the Printed Circuit Board Assembly (PCBA) in the BMU 40 units has a short creepage distance of 1.8 mm between two distinct and proximal areas on the printed circuit board, whereas the IEC standard (60601 1, third edition) requires a creepage distance of 4 mm.
Context
GovInfoNow's records show Maquet Cardiovascular Us Sales, Llc has 14 other recalls on record since 2018-10-25. Other recent recalls include the Maquet HLS Set Advanced 7.0 Recall (2020-11-20) for sterility barrier breaches and the Maquet Yuno II Mobile Operating Table Recall (2020-09-24) for electric shock risk from exposed wiring. Other recent recalls of similar products include the Philips Allura Xper FD10C Medical System Recall (2025-12-15) and the Philips Allura Xper FD10F Medical System Recall (2025-12-15), both citing risk of electrical short-circuits.