Ra Medical Systems, Inc. voluntarily recalled the DABRA Laser (RA-308 Excimer Laser) due to a footswitch hazard that could allow the laser to continue firing if liquids spill on it, according to the FDA. The agency announced the Class II recall on Oct. 16, 2020.
How to tell if you have it
- Brand: Ra Medical Systems
- Product: DABRA Laser (RA-308 Excimer Laser)
- Model Number: RA-308
- Catalog Number: 1200-0001-01
- Serial Numbers: RA00027, RA00046, RA00049, RA00050, RA00054, RA00056, RA00057, RA00058, RA00064, RA00069, RA00073, RA00081, RA00083, RA00084, RA00085, RA00089, RA00097, RA00098, RA00114, RA00115, RA00120, RA00121, RA00125, RA00127, RA00135, RA00138, RA00148, RA00153, RA00154, RA00160, RA00164, RA00165, RA00167, RA00170, RA00175, and RA00177
- Quantity: 36 devices
- Distribution: Sold in AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, WA, and Japan
The hazard
The footswitch does not meet required specifications for protection from ingress of solids or liquids. If liquids spill near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.
Context
GovInfoNow's records show Ra Medical Systems, Inc. has had five other recalls since Feb. 15, 2018. Other recent recalls of similar products include Lutronic DermaV Laser System Recall (Aug. 6, 2026), Lutronic CLARITY II Laser System Model 1110200210 Sparks and May Burn Patients (Jan. 27, 2026), and Arrow Medical Drainage Kits Recall (April 21, 2026).