Lower risk Medical device recall · FDA · Published October 26, 2020

Siemens Atellica IM 130: False test results from custom settings resetting

Siemens recalls Atellica IM 1300 analyzers with software version V1.23.1 or lower that may reset custom settings causing false test results. Affected units have Siemens Material Number 11066001.

🇺🇸 FDA recall #Z-0473-2021Medical deviceSiemens Healthcare ›
Take action

Contact the company

Recalling firmSiemens Healthcare Diagnostics, Inc.511 Benedict Ave · Tarrytown, NY 10591-5005
✎ GovInfoNow plain-language explainer

Siemens is recalling Atellica IM 1300 analyzers with software version V1.23.1 or lower. These devices may reset custom test settings when scanning new reagent kits, leading to inaccurate test results. This could affect medical testing accuracy for clinical specimens. Read more at govinfonow.com/recall/64053-siemens-atellica-im-130-false-test-results-from-custom-settings.

Source Summary by GovInfoNow · govinfonow.com/recall/64053-siemens-atellica-im-130-false-test-results-from-custom-settings · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0473-2021

Cite this page

“Siemens Atellica IM 130: False test results from custom settings resetting.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/64053-siemens-atellica-im-130-false-test-results-from-custom-settings. Based on FDA recall #Z-0473-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0473-2021%22&limit=1).

Siemens Healthcare Diagnostics, Inc. recalled Atellica IM 1300 Analyzers on Oct. 26, 2020, according to the FDA. The hazard involves test definition scanning resetting custom settings, causing falsely low or high results.

How to tell if you have it

  • Brand: Siemens Healthcare
  • Product: Atellica IM 1300 Analyzer
  • Siemens Material Number (SMN): 11066001
  • Software version: V1.23.1 or lower
  • UDI: 00630414007960

The hazard

Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values.

Context

GovInfoNow's records show Siemens Healthcare Diagnostics, Inc. has 150 other recalls on record since 2026-08-17. Other recent recalls of similar products include Beckman Coulter DxC 500i Modules with Software 1.5 Risk: Erroneous Test Results (2026-07-15) and Roche Elecsys Anti-TSHR Assay Recall: Inconsistent Results Risk Incorrect Diagnosis (2025-11-21).

How to tell if you have this one

Check for Siemens Material Number 11066001 and software version V1.23.1 or lower. The device is an automated immunoassay analyzer for clinical specimen testing.

  • Brand
    Siemens Healthcare
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    Worldwide distribution - US Nationwide distribution and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Peru, Poland, Portugal, Qatar Republic, Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay, Vatikancity, Vietnam.
Recall numberProduct, codes & lotsQuantity
Z-0473-2021Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001; Software version: V1.23.1 or lower UDI: 00630414007960US 392 units; OUS 449 units
Z-0474-2021Atellica IM 1600 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066000; Software version: V1.23.1 or lower UDI: 00630414002026US: 214 units; OUS: 1082 units
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version is V1.23.1 or lower by checking the device's settings menu.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide distribution and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 3.5 years after the firm started it.

1. Firm started the recall: Oct 26, 202012. FDA classified it Class II: Nov 19, 202023. Published in FDA’s Enforcement Report: Nov 25, 202034. FDA terminated the recall: Apr 19, 2024420202024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0473-2021 (October 26, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0473-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA