The FDA announced on 2020-11-19 that Steris Corporation is recalling Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument due to a risk of bacterial contamination. GovInfoNow's records show the recall was initiated voluntarily by the firm.
How to tell if you have it
- Brand: Steris
- Product: Medisafe Distal Duck Kit and Duck Bag, Cleaner, ultrasonic, medical instrument
- Model# M20350 Lot# 1811273,1811277,1906484,1908593,1909635,1910651, 1911658,2001730,2004770
- Model# M20358 Lot# 1903408,1906508,1907556,1909644,2001718,2004772
- Model# M20359 Lot# 1811278,1906499,1910649,1911680,1911691,1912708, 2001719,2004768
- Model# M20400 Lot# 1901340,1901349,1902374,1904448,1904464,1905505, 1906529,1907571,1908595,1909624,1910653,1912700, 2001717,2004766,2004774,2006797
The hazard
The FDA said certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. GovInfoNow's records show there is an improbable risk to users from exposure to this bacteria, and no risk to patients.
Context
GovInfoNow's records show Steris Corporation has 31 other recalls on record since 2017-06-15. Recent Steris recalls include a software issue for the HexaVue Focus Integration System (2026-09-23) and a challenge pack recall (2026-08-31). Other recent recalls of similar products include Medline Centurion Shave Kits Recall: Non-sterile Alcohol Prep Pads Risk Infection (2026-03-16), Medline Convenience Kits Recall: Risk of Microbial Contamination from Packaging Wrinkles (2026-06-12), and Medline Convenience Kits Recall: Fungal Contamination Risk (2026-08-11).