Use caution Medical device recall · FDA · Published November 20, 2020

Haag-Streit Surgical Floor Stands with Software Error Risk Patient Eye Injury

Haag-Streit recalls surgical floor stands FS 2-11 and FS 2-15 with specific software versions that may cause eye injury during use in ophthalmology. Affected units have serial numbers starting from 1001.

🇺🇸 FDA recall #Z-0697-2021Medical deviceHaag-Streit ›
Take action

Contact the company

Recalling firmHaag-Streit USA Inc3535 Kings Mills Rd · Mason, OH 45040-2303
✎ GovInfoNow plain-language explainer

Haag-Streit is recalling surgical floor stands FS 2-11 and FS 2-15 that come with certain software versions. These stands could potentially cause eye injury to patients if the focusing mechanism doesn't stop when the button is released during surgery. Read more at govinfonow.com/recall/63831-haag-streit-surgical-floor-stands-with-software-error-risk-patient.

Source Summary by GovInfoNow · govinfonow.com/recall/63831-haag-streit-surgical-floor-stands-with-software-error-risk-patient · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0697-2021

Cite this page

“Haag-Streit Surgical Floor Stands with Software Error Risk Patient Eye Injury.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63831-haag-streit-surgical-floor-stands-with-software-error-risk-patient. Based on FDA recall #Z-0697-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0697-2021%22&limit=1).

Haag-Streit USA Inc recalled Haag Streit Surgical Floor stands FS 2-11 and FS 2-15 used with specific operating microscopes due to a software error that may cause focusing movement to continue after the footswitch is released, posing a risk to patients' eyes. The Food and Drug Administration announced the Class II recall on November 20, 2020.

How to tell if you have it

  • Haag-Streit FS 2-11 (Ref no. 615H511), Serial Numbers from 1001
  • Haag-Streit FS 2-15 (Ref no. 615H515), Serial Numbers from 1001
  • HS Hi-R NEO 900 (Ref no. 657 820), Serial Numbers from 101
  • HS Hi-R NEO 900A (Ref no. 657 821), Serial Numbers from 101
  • Software REF 615 588 versions 2.0 to 3.3

The hazard

The software error in REF 615 588 versions 2.0 to 3.3 may cause the focusing movement to not stop when releasing the footswitch button, resulting in a potential risk to the patient's eye.

Context

GovInfoNow's records show Haag-Streit USA Inc has issued three other recalls since July 31, 2019, including a November 3, 2021 recall for OCT-Camera 211 01 A1 software due to laser safety risks and an April 15, 2021 recall for Eyesuite 9.3.1 software regarding patient data mix-up risks. Other recent recalls of similar products include a November 7, 2025 Da Vinci Surgical System recall for software errors causing instrument failure, a February 23, 2026 Stryker CHROMOPHARE Softlit Ring Surgical Light recall for ceiling support risks, and an November 12, 2025 Howmedica Stryker CranialMask Tracker recall for software errors preventing device use.

How to tell if you have this one

Check if your Haag-Streit surgical floor stand has serial numbers starting from 1001. The affected models are FS 2-11 (Ref 615H511) and FS 2-15 (Ref 615H515) with software versions 2.0 to 3.3. Also check the unique device identifier codes on the product.

  • Brand
    Haag-Streit
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify the software version on your Haag-Streit surgical floor stand. If it is 2.0 to 3.3, it is affected.

  2. Contact Manufacturer

    Reach out to Haag-Streit USA Inc for guidance on next steps and potential remedies.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution in the state of OK.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 40 days after the firm started it.

1. Firm started the recall: Nov 20, 202012. FDA classified it Class II: Dec 23, 202023. Published in FDA’s Enforcement Report: Dec 30, 202032020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0697-2021 (November 20, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0697-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA