The FDA announced on 2020-11-20 that Ortho-Clinical Diagnostics, Inc. is recalling the enGen Laboratory Automation System and enGen Track System due to a software anomaly that may cause samples to be mislabeled. This error could lead to results being associated with the wrong patient sample and potentially result in inappropriate intervention with the potential for injury to the patient.
How to tell if you have it
- Brand: Ortho-Clinical
- Product: enGen Laboratory Automation System, enGen Track System
- Catalog Number: ENGEN
- Configuration: VITROS Automation Solutions Using All TCAutomationTM configured with Thermo Scientific Aliquoter & Labeller Module
- Software Versions: 4.6 and below
- Units affected: 40 units
The hazard
A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample, according to the FDA. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.
Context
GovInfoNow's records show Ortho-Clinical Diagnostics, Inc. has 53 other recalls on record since 2017-07-10. Recent brand recalls include Ortho-Clinical LAC Slides Recall (2025-07-29), Ortho-Clinical VITROS Ca Slides Recall (2025-07-09), and Ortho-Clinical VITROS OP Reagent Gen 15 May Cause Inaccurate Test Results (2025-06-12). Other recent recalls of similar products include GE Medical Systems ECG Systems Software Error Risks Patient Data Mix-Up (2026-04-21), Philips IntraSight Plus Software Error May Cause Wrong Patient Diagnosis (2026-08-06), and Da Vinci Surgical System: Software Error May Cause Instrument Failure (2025-11-07).