B Braun Medical, Inc. recalled the APEX Compounding System Control Panel Module on Nov. 23, 2020, according to the FDA. The device may fail to alert users to air bubbles exceeding 3% of the ordered volume. This could result in compounded drugs that do not meet specified accuracy ranges.
How to tell if you have it
- Brand: B Braun
- Product: APEX Compounding System Control Panel Module
- REF: 601237
- Product Code: AX1000
- Software Version: 1.0.5.57 or earlier
- Units affected: 85
The hazard
The compounding system has the potential to not immediately interrupt compounding and alert the user to the presence of an air bubble exceeding 3% of the ordered volume for an ingredient. This could result in a compounded drug that does not meet the specified accuracy range for an individual ingredient.
Context
GovInfoNow's records show B Braun Medical Inc has 66 other recalls on record since Oct. 15, 2013. Other recent recalls of similar products include B. Braun Medical 100mL Sodium Chloride Injection Recall (Aug. 19, 2026), B. Braun Medical Lactated Ringer's Injection Recall (Aug. 12, 2026), and B Braun Discofix Four Way Stopcock Recall (Aug. 4, 2026). Other recent recalls of similar products include Siemens Healthcare Syva EMIT 2000 Quinidine Assay Recall (July 13, 2026), Broselow Domestic Complete ALS Organizer Software Recall (Dec. 15, 2025), and Beckman Coulter Emit 2000 Quinidine Assay Recall (July 13, 2026).