The FDA announced on November 30, 2020, that Siemens Medical Solutions USA, Inc. is recalling Artis zee, zeego, Q, Q.zen, and Pheno medical imaging devices because the measured Air Kerma Rate can exceed regulatory limits during specific Fluoro Override operations. Owners should contact Siemens for remedy details.
How to tell if you have it
- Brand: Siemens
- Product names: Artis zee, zeego, Q, Q.zen, Pheno
- Model numbers: 10094135, 10094137, 10094139, 10094141, 10280959, 10849000, 10848280, 10848281, 10848282, 10848283, 10848353, 10848354, 10848355
- Quantity: 5602 units
The hazard
According to the FDA, when users change frame rates from lower frequency (e.g., 10 f/s) to higher frequency (e.g., 30 f/s) during continuous release of Fluoro "Fluoro Override," the measured Air Kerma Rate can exceed regulatory limits.
Context
GovInfoNow's records show Siemens Medical Solutions USA, Inc. has 182 other recalls on record since June 6, 2017. Other recent recalls of similar products include Siemens ARTIS icono Ceiling Monitor Recall (June 8, 2026), Siemens ARTIS icono ceiling recall (May 20, 2026), and Siemens ARTIS Pheno VE30A/VE40A recall (March 12, 2026). Other recent recalls of similar products include AGFA DR 800 X-Ray System Recall (November 18, 2025), Philips Brilliance CT 64 Medical Scanner Recall (August 7, 2026), and Canon Alphenix INFX-8000H X-ray System Recall (December 19, 2025).