The FDA announced on Dec. 2, 2020, that Beckman Coulter Inc. is recalling MicroScan Pos MIC Panel Type 34 due to a potential for false elevated daptomycin susceptibility results. The agency classified the recall as Class II.
How to tell if you have it
- Brand: Beckman Coulter
- Product: MicroScan Pos MIC Panel Type 34
- UPC: 15099590657550
- Lot Number: 2021-06-10
- Catalog Number: B1017-216
- Quantity: 685 panel boxes (13,700 panels)
The hazard
The recall was initiated after a customer report and internal investigation confirmed that a portion of the affected product lot failed quality control for ATCC E. Faecalis 29212 versus daptomycin. This failure creates a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.
Context
GovInfoNow's records show Beckman Coulter, Inc. has 133 other recalls on record since 2017-06-20. Other recent recalls of similar products include DxC 500i Clinical Analyzer Modules May Show Incorrect Test Results Due to Software Issue (2026-09-02), Beckman Coulter DxC 500 AU Modules with Software 1.5 Risk Patient Test Errors (2026-07-15), and Beckman Coulter MicroScan LabPro V5.03 Panel Update 11 DVD Kit: Incorrect Algorithms May Affect Test Results (2026-06-26). Additional similar recalls include Beckman Coulter Emit 2000 Quinidine Assay Recall: Risk of Incorrect Patient Treatment (2026-07-13), Beckman Coulter REMISOL Advance Software Issue: Incorrect Test Results Risk (2026-04-29), and Beckman Coulter DxC 700 AU Recall: Delay in Results May Cause Diagnosis Delays (2026-01-30).