GE NM/CT 870 CZT System Recall: Loose Rotor Bearing Risk

GE Healthcare recalls NM/CT 870 CZT system model H3906CW due to loose rotor bearing screws that may cause a bearing to release during use. Affected units have specific serial numbers listed.

🇺🇸 FDA recall #Z-0855-2021Medical deviceGE Healthcare ›
Take action

Contact the company

Recalling firmGE Healthcare, LLC3000 N Grandview Blvd · Waukesha, WI 53188-1615
✎ GovInfoNow plain-language explainer

GE Healthcare is recalling certain NM/CT 870 CZT systems because loose rotor bearing screws could cause one of four bearings to release. This poses a risk of equipment malfunction during nuclear medicine imaging procedures. Read more at govinfonow.com/recall/63679-ge-nm-ct-870-czt-system-recall-loose-rotor-bearing.

Source Summary by GovInfoNow · govinfonow.com/recall/63679-ge-nm-ct-870-czt-system-recall-loose-rotor-bearing · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0855-2021

Cite this page

“GE NM/CT 870 CZT System Recall: Loose Rotor Bearing Risk.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63679-ge-nm-ct-870-czt-system-recall-loose-rotor-bearing. Based on FDA recall #Z-0855-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0855-2021%22&limit=1).

The FDA announced on Dec. 10, 2020, that GE Healthcare LLC is recalling its GE NM/CT 870 CZT system Model Number H3906CW because loose rotor bearing screws may cause a rotor bearing to release from the detector. The agency classified the recall as Class II.

How to tell if you have it

  • Brand: GE Healthcare
  • Product: GE NM/CT 870 CZT system
  • Model Number: H3906CW
  • Serial numbers: 87CY63028, 87CZ63044, 87CZ63045, 87CZ63042, 87CY63025, 87CY63029, 87CY63027, 87CY63030, 87CZ63046, 87CY63034, 87CY63032, 87CY63033
  • Quantity: 12 units

The hazard

GE Healthcare stated that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.

Context

GovInfoNow's records show GE Healthcare LLC has 97 other recalls on record since Oct. 4, 2017. Recent brand recalls include SIGNA Premier MRI Scanners with steel fittings risk (Feb. 27, 2026), Centricity Universal Viewer inaccurate measurements (July 15, 2022), and SIGNA Premier MRI Scanner hearing loss risk (June 24, 2022). Other recent recalls of similar products include Philips Spectral CT Recall (Jan. 7, 2026), Philips Brilliance iCT CT Scanner Recall (Jan. 7, 2026), and Philips IQon Spectral CT Recall (Jan. 7, 2026).

How to tell if you have this one

Check for the model number H3906CW on the system. Affected units have serial numbers listed in the recall notice: 87CY63028, 87CZ63044, 87CZ63045, 87CZ63042, 87CY63025, 87CY63029, 87CY63027, 87CY63030, 87CZ63046, 87CY63034, 87CY63032, 87CY63033.

  • Brand
    GE Healthcare
  • Category
    Medical device
FDA recall event · 6 products

All 6 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide
Recall numberProduct, codes & lotsQuantity
Z-0853-2021GE NM/CT 850 system Model Number H3907AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.; Serial numbers: 85EY67048, 850Y61018, 85EY67045, 85EY67046, 85EY67047, 850Z61038, 85EY67043, 850Z61042, 85EZ67056, 850Z61043, 850Z61044, 850Z61045, 850Z61041, 850Z61039, 850Z61040, 85EZ67055, 85EZ67053, 85EZ67054, 850Y61019, 85EY67044, 85EY67049, 85EZ67052, 85EY67050, 85EZ6705124 units
Z-0854-2021GE NM/CT 860 system Model Number H3908AD - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.; Serial numbers: 860Y62016, 860Y62019, 860Y62017, 860Z62032, 860Y62018, 860Y62015, 860Y62012, 860Y62020, 860Y62021, 860Y62013, 860Z62033, 860Z6203112 units
Z-0855-2021GE NM/CT 870 CZT system Model Number H3906CW - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.; Serial numbers: 87CY63028, 87CZ63044, 87CZ63045, 87CZ63042, 87CY63025, 87CY63029, 87CY63027, 87CY63030, 87CZ63046, 87CY63034, 87CY63032, 87CY6303312 units
Z-0856-2021Discovery NM 630 (Model Number H3101RH), NM 830 (Model Number H3910AC), 830 NM Cyber Defense (Model Number H3910AG) - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.; Discovery NM 630 (Model Number H3101RH), Serial numbers: 63GY50110, 63GY50119, 63GY50116, 63GZ50141, 63GY50120, 63GY50108, 63GY50121, 63GZ50142, 63GY50112, 63GZ50139, 63GY50107, 63GZ50138 NM 830 (Model Number H3910AC), Serial Numbers: 830Z60199, 830Z60182, 830Z60183, 830Z60192, 830Y60089, 830Y60105, 830Y60079, 830Z60207, 830Y60092, 830Y60102, 830Y60081, 830Z60200, 830Y60106, 830Z60201, 83EZ88001, 830Y60111, 830Y60094, 830Z60211, 830Y60101, 830Z60210, 830Y60118, 830Z60195, 830Z60194, 830Z60198, 830Z60181, 830Z60202, 830Y60097, 830Z60196, 830Z60180, 830Y60078, 830Y60115, 830Y60114, 830Y60108, 830Y60100, 830Y60087, 830Y60084, 830Z60188, 830Z60206, 830Y60112, 830Y60082, 830Y60093, 830Z60191, 830Y60107, 830Y60080, 830Z60190, 830Y60103, 830Z60193, 830Y60095, 830Y60119, 830Z60209, 830Z60205, 830Y60073, 830Y60120, 830Y60083, 830Y60116, 830Z60203, 830Y60085, 830Y60077, 830Z60184, 830Y60086, 830Z60187, 830Y60113, 830Z60208, 830Y60110, 830Z60204, 830Y60072, 830Z60179, 830Z60186 830 NM Cyber Defense (Model Number H3910AG), Serial Numbers: 830Z6018581 units
Z-0857-2021Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.; Discovery 670 DR (Model H3100BT), Serial Numbers: DRGY56033, DRGY56037, DRGY56035, DRGY56030, DRGZ56043, DRGZ56042, DRGY56032, DRGY56031, DRGY56034, DRGY56036 NM/CT 870 DR (Model 3100Q), Serial Numbers: 870Y64066, and (Model H3100AS), Serial Numbers: 870Y64077, 870Y64064, 870Y64081, 870Y64079, 870Y64067, 870Y64074, 870Z64117, 870Y64078, 870Y64080, 870Y64069, 870Y64085, 870Z64121, 870Z64119, 870Z64122, 870Y64056, 870Y64063, 870Z64114, 870Z64115, 870Y64068, 870Y64082, 870Z64116, 870Z64120, 870Y64065, 870Y64084, 870Y64061, 870Y64083, 870Z64113, 870Y64070, 870Z64123, 870Y64058, 870Z64118, 870Y64057 Discovery NM670 Pro (Model 3100XB), Serial Numbers: PRGY52027, PRGY52031, PRGZ52047, PRGZ52046, PRGY52028, PRGZ52048, PRGY52029, PRGY52030 Discovery 670 ES (Model H3100XF), Serial Numbers: ESGZ5300552 untits
Z-0858-2021Optima NM/CT 640, Model number 3101RL - Product Usage: intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.; Serial Numbers: 64GY51017, 64GY51020, 64GY51022, 64GZ51028, 64GZ51030, 64GZ51029, 64GY510217 units
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Numbers

    Compare your system's serial number with the list provided in the recall notice: 87CY63028, 87CZ63044, 87CZ63045, 87CZ63042, 87CY63025, 87CY63029, 87CY63027, 87CY63030, 87CZ63046, 87CY63034, 87CY63032, 87CY63033.

How this recall unfolded

Recall timeline

FDA published this recall 41 days after the firm started it.

1. Firm started the recall: Dec 10, 202012. FDA classified it Class II: Jan 13, 202123. Published in FDA’s Enforcement Report: Jan 20, 2021320202021
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0855-2021 (December 10, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0855-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA