Skip to content
Use caution Medical device recall · FDA · Published December 15, 2020

Penumbra JET 7 Kit: Distal Tip Damage Risk During Use

Penumbra recalls JET 7 Reperfusion Catheter kits due to potential distal tip damage during use that could cause vessel injury or death if contrast injection pressure is applied.

🇺🇸 FDA recall #Z-0849-2021Medical devicePenumbra ›
Take action

Contact the company

Recalling firmPenumbra Inc.1 Penumbra · Alameda, CA 94502-7610
✎ GovInfoNow plain-language explainer

Penumbra is recalling its JET 7 Reperfusion Catheter kits because the catheter tip may get damaged during use. This damage could cause vessel injury or death if contrast injection pressure is applied. Affected units were sold between specific lot numbers and dates. Read more at govinfonow.com/recall/63656-penumbra-jet-7-kit-distal-tip-damage-risk-during-use.

Source Summary by GovInfoNow · govinfonow.com/recall/63656-penumbra-jet-7-kit-distal-tip-damage-risk-during-use · Updated August 19, 2026 · FDA enforcement record: FDA #Z-0849-2021

Cite this page

“Penumbra JET 7 Kit: Distal Tip Damage Risk During Use.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/63656-penumbra-jet-7-kit-distal-tip-damage-risk-during-use. Based on FDA recall #Z-0849-2021 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0849-2021%22&limit=1).

Penumbra Inc. recalled Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7KIT due to potential vessel damage and patient injury or death, the FDA said on 2020-12-15.

How to tell if you have it

  • Brand: Penumbra
  • Product Name: Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing
  • Catalog Number: 5MAXJET7KIT
  • UDI Code: 00815948020962
  • Lot Number Range: C15441 thru C19176 F88882 thru F100278

The hazard

The catheter may become susceptible to distal tip damage during use. Distal tip damage in conjunction with pressurization of contrast injection may result in potential vessel damage, and subsequent patient injury or death.

Context

GovInfoNow's records show Penumbra Inc. has issued three other recalls since 2017-06-09, including the 2026-07-24 Penumbra RED 62 Reperfusion Catheter Recall for fracture risk and the 2026-06-03 Penumbra RED 68 Reperfusion Catheter Recall for fracture risk. Other recent recalls of similar products include the 2026-08-24 Boston Scientific Imager II Angiographic Catheters Recall for tip fracture risk and the 2026-06-15 Argon Medical L-Cath PICC Catheters Recall for leaking or cracking.

How to tell if you have this one

Check for the lot number range C15441 to C19176 and the F88882 to F100278 on the packaging. The product is sold under the catalog number 5MAXJET7KIT.

  • Brand
    Penumbra
  • Category
    Medical device
FDA recall event · 4 products

All 4 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class I
  • Status
    Terminated
  • Distribution
    Worldwide distribution: U.S. (nationwide) including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO , MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. And O.U.S (countries) of: Canada, Chile, Colombia, Costa Rica, Dominican Republic, Hong Kong, Israel, Japan, Korea, Lebanon, Malaysia, Nepal, New Zealand, Oman, Panama, Saudi Arabia, Singapore, South Africa, Taiwan, Thailand, United Arab Emirates, and Vietnam.
Recall numberProduct, codes & lotsQuantity
Z-0848-2021Penumbra JET 7 Reperfusion Catheter REF 5MAXJET7 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.; Product Name: Penumbra JET 7 Reperfusion Catheter Catalog Number: 5MAXJET7 UDI Code: 00815948020955 Lot Number Range: C15558 thru C17703 F87621 thru F101011 H108513,368 catheters
Z-0849-2021Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.; Product Name: Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing Catalog Number: 5MAXJET7KIT UDI Code: 00815948020962 Lot Number Range: C15441 thru C19176 F88882 thru F10027827,499 catheters
Z-0850-2021Penumbra JET 7 KIT Reperfusion Catheter+Penumbra Hi-Flow Aspiration Tubing REF 5MAXJET7L138KIT The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.; Product Name: Penumbra JET 7 Reperfusion Catheter and Penumbra Hi-Flow Aspiration Tubing Catalog Number: 5MAXJET7L138KIT UDI Code: 00815948021594 Lot Number Range: F97123 thru F98981 Note - none were distributed0 catheters
Z-0851-2021Penumbra JET 7MAX KIT Penumbra JET 7 Reperfusion Catheter+ MAX Delivery Device + Hi-Flow Aspiration Tubing REF J7XKIT132 The Penumbra JET 7 Reperfusion Catheter with Xtra Flex technology (JET 7 Xtra Flex) is a component of the Penumbra system. JET 7 Xtra Flex delivers aspiration from the Aspiration Pump to the site of occlusion to assist in the efficient removal of thrombus from the neurovasculature. The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.; Product Name: JET 7MAX (Penumbra JET 7 Xtra Flex + MAX Delivery Device) Catalog Number: J7XKIT132 UDI Code: 00815948021273 Lot Number Range: C21487 thru C21772 F97946 thru F9849715 catheters
Based on the official FDA remedy

What to do — step by step

  1. Check for Damage

    Inspect the catheter tip for signs of damage before use.

  2. Contact Manufacturer

    Reach out to Penumbra Inc. for specific guidance on safe use or replacement options.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 47 states.

Alaska: named in the noticeAKMaine: named in the noticeMEVermont: nationwideVTNew Hampshire: named in the noticeNHWashington: named in the noticeWAIdaho: named in the noticeIDMontana: named in the noticeMTNorth Dakota: named in the noticeNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: named in the noticeMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: nationwideORNevada: named in the noticeNVWyoming: named in the noticeWYSouth Dakota: named in the noticeSDIowa: named in the noticeIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: named in the noticeCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: named in the noticeKYWest Virginia: named in the noticeWVVirginia: named in the noticeVAMaryland: named in the noticeMDDelaware: named in the noticeDEArizona: named in the noticeAZNew Mexico: named in the noticeNMKansas: named in the noticeKSArkansas: named in the noticeARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: named in the noticeSCDistrict of Columbia: named in the noticeDCOklahoma: named in the noticeOKLouisiana: named in the noticeLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: named in the noticeHITexas: named in the noticeTXFlorida: named in the noticeFLPuerto Rico: named in the noticePR

FDA distribution: “Worldwide distribution: U.S. (nationwide) including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO , MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, Puerto Rico, SC, SD, TN,…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Pennsylvania, Puerto Rico, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming.

Recall timeline

FDA terminated this recall 3.4 years after the firm started it.

1. Firm started the recall: Dec 15, 202012. FDA classified it Class I: Jan 18, 202123. Published in FDA’s Enforcement Report: Jan 27, 202134. FDA terminated the recall: May 14, 2024420202024
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0849-2021 (December 15, 2020) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0849-2021 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA