Abbott Point Of Care Inc. recalled Abbott i-STAT CG8+ cartridges on Dec. 22, 2020, because ionized calcium tests may show higher than expected results when stored at room temperature for too long, the FDA said. Owners should contact the company for a refund.
How to tell if you have it
- Brand: Abbott
- Product: i-STAT CG8+ cartridges
- Lot numbers: *20100 to *20339 and *20340 and above (where * = A, K, L, M, N, W or Y)
- GTIN: 00054749000163
The hazard
The hazard is that ionized calcium tests on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above.
Context
GovInfoNow's records show Abbott Point Of Care Inc. has had 8 other recalls on record since Sept. 12, 2017. Other recent recalls include Abbott i-STAT G3+ Cartridges May Show Wrong Blood Gas Results (April 1, 2026), Abbott i-STAT CG4+ Cartridge Recall: Lack of Premarket Clearance (Aug. 21, 2025) and Abbott i-STAT EG7+ Cartridge Recall: Lack of Premarket Clearance (Aug. 21, 2025).