The FDA announced on Jan. 7, 2021, that Becton Dickinson & Co. is recalling BioGX SARS-CoV-2 Reagents for BD MAX System catalog number 444213 due to potential mislabeled master mix pouches. This error could lead to false negative SARS-CoV-2 test results.
How to tell if you have it
- Brand: Becton Dickinson
- Product: BioGX SARS-CoV-2 Reagents for BD MAX System
- Catalog number (REF): 444213
- UPC: 00382904442132
- Lot numbers: K20-507, K20-508, K20-509, K20-516, K20-517, K21-020, K21-021
The hazard
Some N1 and N2 master mix pouches may be mislabeled, resulting in kits containing two pouches of the same type instead of one of each. If laboratory personnel use two identical pouches, there is a low potential for a false negative SARS-CoV-2 result, which could lead to increased disease progression risk and additional spread of the virus.
Context
GovInfoNow's records show Becton Dickinson & Co. has 43 other recalls on record since 2017-06-01. Recent brand recalls include a BD GasP CO2 Pouch System recall on 2026-04-07 and a BD Kiestra Isolate Suspension Cuvette Array recall on 2025-10-23. Other recent recalls of similar products include a Biomerieux VITEK 2 AST-N461 Test Kit Recall (2026-07-24) and a BioFire Respiratory Panel 2.1 Recall (2025-10-22).